MINIMED 770G OUS SYSTEM BLE CONNECT 3.0 MMOL/L
Report
- Report Number
- 2032227-2023-169686
- Event Type
- Malfunction
- Date Received
- March 13, 2023
- Date of Event
- February 18, 2023
- Report Date
- May 1, 2023
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000456634
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- 003
Narratives
P-CAP LOCKED IN PLACE PROPERLY DURING TESTING. PROCEED IT BY USING A NEW BATTERY CAP AND A NEW BATTERY. UNIT POWER UP PROPERLY AFTER BATTERY INSTALLATION. UNIT WAS DOWNLOADED SUCCESSFULLY USING (THUMP SOFTWARE) FOR REFERENCE. THE ADAPT TOOL WAS UTILIZED TO SEARCH FOR ANY ALARMS THAT MAY HAVE OCCURRED IN THE PAST OR AROUND THE COMPLAINT DATE THAT MIGHT OF TRIGGER THE REASON COMPLAIN. THE ADAPT TOOL RECORDED LOW BATTERY ALERT ON 02/01/2023, 02/04//2023, 02/08/2023, 02/10/2023, 02/13/2023, 02/16/2023, 02/19/2023 AND ON 02/22/2023. PROCEEDED WITH THE FOLLOWING TESTING TO ASSURE PROPER FUNCTIONALITY OF THE UNIT. UNIT PASSED THE SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT AND DISPLACEMENT TEST. NO LOW BATTERY ALARMS OR UNEXPECTED BATTERY POWER LOSS NOTED DURING TESTING. UNIT FUNCTIONING PROPERLY. THE POWER MANAGEMENT PARAMETERS GRAPH CONFIRMED THE UNLOADED VOLTAGE (UNLOADED VLITH) AND LOADED VOLTAGE (LOADED VLITH) WAS WITHIN SPEC RANGE. PUMP RECEIVED WITH NORMAL OPERATING CURRENTS. HOWEVER, SEVERELY CRACKED CASE ON THE BATTERY COMPARTMENT SIDE, CRACKED BATTERY HOUSING AND CRACKED BATTERY TUBE THREADS THAT MIGHT OF CAUSE THE BATTERY NOT MAKING CONTACT WITH BATTERY CAP. PER VISUAL INSPECTION ALL CONNECTORS INCLUDING BATTERY FLEX CONNECTOR WERE PLUGGED IN PROPERLY. NO MOISTURE DAMAGE NOTED DURING VISUAL INSPECTION. UNIT ALSO RECEIVED WITH PEELING SERIAL NUMBER LABEL, CRACKED KEYPAD AT SELECT BUTTON AND SCRATCHED CASE. IN CONCLUSION NO LOW BATTERY ALARMS NOTED DURING TESTING. HOWEVER, SEVERAL LOW BATTERY ALERTS WERE RECORDED. UNIT MAY HAVE HAD AN INTERMITTENT FAILURE NOT DETECTED DURING OUR TESTING. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE NGP 770G INSULIN PUMP WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE NGP INSULIN PUMP, WHICH IS MARKETED IN THE UNITED STATES. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED A LOW BATTERY ALERT. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLESHOOTING WAS PERFORMED. THE CUSTOMER WILL DISCONTINUE USING THE PUMP. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 515076 | MINIMED 770G OUS SYSTEM BLE CONNECT 3.0 MMOL/L | AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1881 | HG5KTJTZZ | 000000763000456634 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |