ELECTRODE ONESTEP CPR A/A GC, SINGLE
Report
- Report Number
- 1218058-2023-00056
- Event Type
- Death
- Date Received
- March 10, 2023
- Date of Event
- March 2, 2023
- Report Date
- March 2, 2023
- Manufacturer
- BIO-DETEK INCORPORATED
- Product Code
- MKJ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THE ELECTRODES WERE RETURNED TO ZOLL MEDICAL UNITED KINGDOM FOR EVALUATION. EVALUATION OF THE ELECTRODES REVEALED THAT THERE WAS AN EXCESSIVE AMOUNT OF HAIR ATTACHED TO THE GEL. THE ELECTRODES PASSED CONTINUITY TESTING. THE LOGS FROM THE EVENT WERE NOT PROVIDED. BASED ON THE CUSTOMER'S DESCRIPTION AND OBSERVED EXCESSIVE HAIR ON THE ELECTRODE, THIS CLAIM WILL BE CLOSED AS POOR COUPLING BETWEEN PATIENT AND ELECTRODE DUE TO POOR PATIENT PREPARATION. R SERIES OPERATOR'S GUIDE, PATIENT SAFETY, STATES: "EXCESSIVE BODY HAIR OR WET, DIAPHORETIC SKIN CAN INHIBIT ELECTRODE COUPLING TO THE SKIN. CLIP EXCESS HAIR AND DRY ANY MOISTURE FROM THE AREA WHERE AN ELECTRODE IS TO BE ATTACHED." ANALYSIS OF REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.
ZOLL MEDICAL CORPORATION HAS NOT RECEIVED THE PRODUCT FOR EVALUATION AND THIS COMPLAINT IS STILL UNDER INVESTIGATION.
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO DEFIBRILLATE A 62-YEAR-OLD MALE PATIENT, A LOUD POP NOISE WAS HEARD ON DELIVERY OF THE THIRD SHOCK, ALONG WITH A WHITE FLASH AND SMOKE WERE SEEN FROM THE ELECTRODE PADS. THE PADS WERE THEN REMOVED TO REPLACE AND UPON REMOVAL, A BURN OF AN UNKNOWN DEGREE WAS SEEN ON THE PATIENT'S CHEST. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THE PATIENT SUBSEQUENTLY EXPIRED, HOWEVER IT WAS NOT A RESULT OF THE REPORTED MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 577235 | ELECTRODE ONESTEP CPR A/A GC, SINGLE | ELECTRODE | MKJ | BIO-DETEK INCORPORATED | 8900-000251-05 | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Male | Death |