WAVEWRITER ALPHA PRIME
Report
- Report Number
- 3006630150-2023-01199
- Event Type
- Injury
- Date Received
- March 10, 2023
- Date of Event
- February 14, 2023
- Report Date
- September 6, 2023
- Manufacturer
- BOSTON SCIENTIFIC NEUROMODULATION
- Product Code
- LGW
- UDI-DI
- 08714729985075
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
SC-1432 (SN B)(6). THE RETURNED IPG WAS ANALYZED AND THE DEVICE EXHIBITED NORMAL CHARACTERISTICS DURING THE VISUAL AND FUNCTIONAL EXAMINATION. WITH ALL THE AVAILABLE INFORMATION, BOSTON SCIENTIFIC CONCLUDES THAT THE REPORTED EVENT COULD NOT BE CONFIRMED.
EVENT DATE: APPROXIMATED BASED ON THE DATE THE MANUFACTURER BECAME AWARE OF THE EVENT. ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED IN THE EVENT: PRODUCT FAMILY: SCS-LINEAR LEADS, UPN: M365SC2352500, MODEL: SC-2352-50, SERIAL: (B)(4), BATCH: 7072689/7073617.
IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING INADEQUATE STIMULATION AND PAIN AT THE IPG SITE. THE PATIENT UNDERWENT A PROCEDURE WHERE IN THE IPG AND LEADS WERE EXPLANTED. ALL EXPLANTED DEVICE COMPONENTS WILL NOT BE RETURNED PER FACILITY POLICY.
IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING INADEQUATE STIMULATION AND PAIN AT THE IPG SITE. THE PATIENT UNDERWENT A PROCEDURE WHEREIN THE IPG AND LEADS WERE EXPLANTED. ALL EXPLANTED DEVICE COMPONENTS WILL NOT BE RETURNED PER FACILITY POLICY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 921090 | WAVEWRITER ALPHA PRIME | STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF | LGW | BOSTON SCIENTIFIC NEUROMODULATION | SC-1432 | 208511 | 08714729985075 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Male | Required Intervention |