KRYPTONITE BONE CEMENT
Report
- Report Number
- 1226001-2010-00005
- Event Type
- Other
- Date Received
- April 5, 2010
- Report Date
- March 22, 2010
- Manufacturer
- DOCTOR'S RESEARCH GROUP, INC.
- Product Code
- GXP
- PMA / PMN Number
- K091382
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
THERE ARE SEVERAL FACTORS THAT MAY HAVE PLAYED A ROLE IN NOT ALLOWING THE KRYPTONITE MATERIAL TO FULLY HARDEN, INCLUDING BUT NOT LIMITED TO; INSUFFICIENT OR IMPROPER SITE PREPARATION, TOO MUCH BLOOD WITHIN SURGICAL SITE, OR IMPROPER MIXING. IT WOULD ALSO APPEAR THAT THE USE OF KRYPTONITE MATERIAL IN THIS SURGICAL PROCEDURE/CONDITION WOULD NOT HAVE BEEN APPROPRIATE. KRYPTONITE BONE CEMENT IS NOT INDICATED FOR USE WERE INTRINSIC STABILITY TO THE BONY STRUCTURE IS REQUIRED. IT WAS ALSO STATED THAT CONTINUED BONE BLEEDING WAS OBSERVED. THE LOT NUMBER WAS NOT AVAILABLE AND THEREFORE, A REVIEW OF THE DEVICE HISTORY RECORDS WERE NOT PERMISSIBLE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, IT WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT. THE IDENTIFIED PRODUCT AND PRODUCT CODE IS DESIGNATED FOR EXPORT ONLY.
A PLIF PROCEDURE WAS PERFORMED AT L4/L5 WHERE KRYPTONITE BONE CEMENT WAS USED. DUE TO A TUMOR, THE SURGEON REMOVED A PORTION OF THE L4 AND L5 VERTEBRAL BODIES AND THE L4/L5 DISC. THE KRYPTONITE BONE CEMENT WAS UTILIZED TO FILL THE VOID AND ACT AS A LOAD SHARING MEDIUM. A PEDICLE SCREW SYSTEM WAS USED TO FURTHER STABILIZE THE SEGMENT. THE AREA WAS DRIED AS MUCH AS POSSIBLE, SOME BONE BLEEDING HOWEVER CONTINUED. UPON SITE CLOSURE, THE KRYPTONITE BONE CEMENT HAD EXPANDED SOMEWHAT BUT WAS STILL WITHIN THE POSTERIOR BORDER OF THE VERTEBRAL BODIES. APPROXIMATELY 10 DAYS POST-OP, THE PATIENT PRESENTED WITH NUMBNESS THROUGH THE BUTTOCK REGION AND A REVISION SURGERY WAS PERFORMED. DURING THE REVISION SURGERY, IT WAS OBSERVED THAT THE KRYPTONITE BONE CEMENT HAD EXPANDED BEYOND EXPECTATION AND PUSHED INTO THE CANAL. DURING REMOVAL, IT WAS NOTED THAT THE KRYPTONITE MATERIAL WAS NOT HARD BUT CLAY LIKE IN CONSISTENCY. THE SURGEON REPORTED THAT THE KRYPTONITE MATERIAL HAD SET ENOUGH TO PROVIDE VERTICAL LOAD BEARING STABILITY BUT WASN'T AS HARD AS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | KRYPTONITE BONE CEMENT | BONE CEMENT | GXP | DOCTOR'S RESEARCH GROUP, INC. | UNKNOWN | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN |