FDA Adverse Event Death Summary report: N

T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY

MDR report key: 16521188 · Received March 10, 2023

Report

Report Number
3013756811-2023-28849
Event Type
Death
Date Received
March 10, 2023
Date of Event
December 21, 2022
Report Date
March 10, 2023
Manufacturer
TANDEM DIABETES CARE
Product Code
QFG
UDI-DI
00850006613793
PMA / PMN Number
K201214
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE PUMP USER GUIDE STATES, "TANDEM DIABETES CARE, INC. RECOMMENDS PERIODICALLY CHECKING THE BATTERY LEVEL INDICATOR, CHARGING THE PUMP FOR A SHORT PERIOD OF TIME EVERY DAY (10 TO 15 MINUTES), AND ALSO AVOIDING FREQUENT FULL DISCHARGES."

Description of Event or Problem · 0

IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ELEVATED BLOOD GLUCOSE (BG) LEVEL AND WAS INCOHERENT. CAUSE OF ELEVATED BG WAS UNKNOWN. REPORTEDLY, THE PUMP BATTERY FULLY DEPLETED AND SUBSEQUENTLY SHUT DOWN DUE TO CUSTOMER NOT CHARGING THE PUMP. THE CUSTOMER¿S SPOUSE DIALED 911, AND PARAMEDICS PERFORMED A FINGERSTICK WHICH INDICATED CUSTOMER¿S BG LEVEL WAS OVER 1,000 MG/DL. CUSTOMER WAS TAKEN VIA AMBULANCE TO THE EMERGENCY ROOM AND SUBSEQUENTLY ADMITTED TO THE HOSPITAL WHERE THEY WERE PLACED IN A MEDICALLY INDUCED COMA. REPORTEDLY, THE CUSTOMER PASSED AWAY ON (B)(6) 2022, TWO DAYS AFTER THE INITIAL EVENT. MULTIPLE FOLLOW-UP ATTEMPTS WERE MADE BUT NO ADDITIONAL INFORMATION WAS AVAILABLE AT THE TIME OF THIS REPORT. THE PUMP WAS NOT UPLOADED TO T:CONNECT, THEREFORE THE PUMP LOG COULD NOT BE INVESTIGATED. IT IS NOT CURRENTLY KNOWN THE EXTENT TO WHICH THE DEVICE SHUTTING OFF DUE TO LOW POWER WAS RELEVANT TO THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1148839 T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY ALTERNATE CONTROLLER ENABLED INFUSION PUMP QFG TANDEM DIABETES CARE 1000354 00850006613793

Patients

Seq Age Sex Outcome Treatment
1 63 YR Male Hospitalization| R| D