FDA Adverse Event Injury Summary report: N

STERIS SYSTEM 1

MDR report key: 1651996 · Received April 6, 2010

Report

Report Number
3000251274-2010-00010
Event Type
Injury
Date Received
April 6, 2010
Date of Event
March 3, 2010
Report Date
April 6, 2010
Manufacturer
STERIS CORPORATION - HOPKINS
Product Code
MED
PMA / PMN Number
K875280
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A STERIS SERVICE TECHNICIAN INSPECTED THE SYSTEM 1 PROCESSOR AND FOUND THE UNIT'S PROCESSING TRAY TO BE FULL OF WATER. THE TECHNICIAN EMPTIED AND CLEANED THE TRAY AND RAN FOUR TEST CYCLES; ALL CYCLES EXHIBITED PASSING RESULTS AND NO ISSUES WERE NOTED WITH THE PROCESSOR LID. AN ADDITIONAL INSPECTION OF THE LID LATCH AND LID ASSEMBLY FOUND THAT THE LID WAS CORRECTLY POSITIONED. AS A PRECAUTION, THE TECHNICIAN REPLACED THE LID LATCH AND LID ASSEMBLY. THE PROCESSOR HAS REMAINED IN USE AND NO ADDITIONAL ISSUES HAVE BEEN REPORTED WITH THE EQUIPMENT. THE HOSPITAL HAS BEEN REMINDED SEVERAL TIMES THAT THE SYSTEM 1 OPERATOR MANUAL REQUIRES THE PROCESSING TRAY AND CHAMBER TO BE CLEANED DAILY WITH A "SOFT CLOTH DAMPENED WITH 70% ISOPROPYL ALCOHOL". NOT FOLLOWING THE DEVICES INDICATIONS FOR USE MAY HAVE CAUSED STRESS ON THE LATCHING MECHANISM, RESULTING IN THE PREMATURE OPENING OF THE LID. REFRESHER IN-SERVICE TRAINING FOR THE HOSPITAL REGARDING PROPER USE OF SYSTEM 1 HAS BEEN OFFERED ON SIX DIFFERENT OCCASIONS, HOWEVER THE HOSPITAL HAS NOT RESPONDED TO STERIS' ATTEMPTS TO SCHEDULE THE TRAINING.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT A SYSTEM 1 PROCESSOR LID OPENED DURING A PROCESSING CYCLE AND S20 STERILANT SPLASHED ONTO A NEARBY HOSPITAL EMPLOYEE. THE EMPLOYEE WAS TREATED WITH BURN DRESSINGS AT THE MINOR INJURIES DEPARTMENT FOR SMALL BLISTERS ON HER LOWER LEGS. THE EMPLOYEE MISSED NO TIME FROM WORK AND REPORTS THAT HER BLISTERS HAVE HEALED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STERIS SYSTEM 1 STERIS SYSTEM 1 MED STERIS CORPORATION - HOPKINS P1000

Patients

Seq Age Sex Outcome Treatment
1 Other