FDA Adverse Event Malfunction Summary report: N

FREESTYLE LITE

MDR report key: 1651994 · Received April 6, 2010

Report

Report Number
2954323-2010-00431
Event Type
Malfunction
Date Received
April 6, 2010
Date of Event
March 10, 2010
Report Date
April 6, 2010
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER'S METER (B) (4) WAS RETURNED AND TESTED WITH RETURNED TEST STRIPS LOT # 0934116. THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION. ADDITIONALLY, THE REPORTED COMPARISON VALUE OF 125 MG/DL WAS FOUND IN THE DEVICE'S INTERNAL MEMORY LOG.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING INACCURATE READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING A READING OF 125 MG/DL COMPARED TO A LAB READING OF 66 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FREESTYLE LITE BLOOD GLUCOSE MONITORING SYSTEM NBW 0934116

Patients

Seq Age Sex Outcome Treatment
1