FDA Adverse Event
Malfunction
Summary report: N
FREESTYLE LITE
MDR report key: 1651823
·
Received April 6, 2010
Report
- Report Number
- 2954323-2010-00428
- Event Type
- Malfunction
- Date Received
- April 6, 2010
- Date of Event
- March 4, 2010
- Report Date
- April 6, 2010
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE METER (B) (4) AND TEST STRIPS (LOT # 0935516) WERE RETURNED FOR AN INVESTIGATION. THE COMPLAINT WAS NOT CONFIRMED. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION DURING CONTROL SOLUTION TESTING. THE READING OF 98 MG/DL WAS FOUND IN THE METER MEMORY LOG ((B) (6) 2010, 4:24 PM). NO REPORT IS REQUIRED.
Description of Event or Problem · 1
A CUSTOMER REPORTED THAT ON (B) (6) 2010 SHE HAD A BLOOD GLUCOSE LAB TEST AT 4:22 PM WITH A RESULT OF 48 MG/DL, AND COMPARED THAT RESULT TO A READING OF 98 MG/DL SHE OBTAINED ON HER BLOOD GLUCOSE METER AT 4:24 PM. THE RESULTS WERE PLOTTED ON A PARKES ERROR GRID AND FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES WAS CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.