FDA Adverse Event Malfunction Summary report: N

BD PHOENIX¿ NID

MDR report key: 16512746 · Received March 9, 2023

Report

Report Number
1119779-2023-00251
Event Type
Malfunction
Date Received
March 9, 2023
Date of Event
February 20, 2023
Report Date
April 7, 2023
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
LQM
UDI-DI
30382904480074
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY THIS COMPLAINT IS FOR MIS-IDENTIFICATION OF ACINETOBACTER BAUMANII AS MORAXELLA CATHATTALIS WHEN USING PHOENIX PANEL NID ((B)(4)) BATCH NUMBER 2110791. THE CUSTOMER DID NOT PROVIDE ISOLATE RETURNS, LAB REPORTS OR PANEL RETURNS FOR THE INVESTIGATION. TO INVESTIGATE, FOUR (4) RETENTION SAMPLES OF THE COMPLAINT BATCH WERE INOCULATED WITH QC ISOLATE ACINETOBACTER BAUMANII 15748 AND EVALUATED IN A PHOENIX M50 FOR IDENTIFICATION RESULTS. ALL FOUR (4) PANELS IDENTIFIED THE ISOLATES AS A. BAUMANII OR A. BAUMANII/CALCOACETICUS COMPLEX, THEREFORE THIS COMPLAINT IS NOT CONFIRMED. A REVIEW OF QUALITY NOTIFICATIONS REVEALED NO QUALITY NOTIFICATIONS FOR THE COMPLAINT BATCH. A REVIEW OF COMPLAINTS REVEALED NO ADDITIONAL COMPLAINTS ON THE COMPLAINT BATCH. COMPLAINT TRENDING WAS PERFORMED AND NO TRENDS WERE IDENTIFIED ASSOCIATED WITH THIS DEFECT. BD ID/AST PLANT QUALITY WILL CONTINUE TO MONITOR FOR TRENDS AND TAKE ACTION AS NECESSARY. H3 OTHER TEXT : SEE H.10

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD PHOENIX¿ NID MISIDENTIFIED ACINETOBACTER BAUMANII AS MORAXELLA CATHATTALIS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ACINETOBACTER BAUMANII WAS IDENTIFIED AS MORAXELLA CATHATTALIS NO IMPACT ON THE PATIENT

Description of Event or Problem · 0

IT WAS REPORTED THAT BD PHOENIX¿ NID MISIDENTIFIED ACINETOBACTER BAUMANII AS MORAXELLA CATHATTALIS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ACINETOBACTER BAUMANII WAS IDENTIFIED AS MORAXELLA CATHATTALIS. NO IMPACT ON THE PATIENT.