FDA Adverse Event Malfunction Summary report: N

MECTALIF ANTERIOR LAG PLATE FLUSH H12 MM

MDR report key: 16511523 · Received March 9, 2023

Report

Report Number
3005180920-2023-00193
Event Type
Malfunction
Date Received
March 9, 2023
Date of Event
February 6, 2023
Report Date
March 9, 2023
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
OVD
UDI-DI
07630345732293
PMA / PMN Number
K221545
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 08-MAR-2023 LOT 2220027: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 21-OCT-2022. EXPIRATION DATE: 2027-08-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH ANOTHER REPORTED EVENT ON THE SAME LOT. NOTE: RECALL IS ON GOING.

Additional Manufacturer Narrative · 0

VISUAL INSPECTION PERFORMED BY R&D PROJECT MANAGER REF. (B)(4) MECTALIF ANT - LAG PLATE FLUSH H12 MM LOT. 2220027. THE FIXING SCREW OF THE COVER PLATE IS BROKEN. ANALYZING THE BREAKAGE SURFACE WITH A MICROSCOPE AND COMPARING IT WITH THE FRACTURE SURFACE OF A TESTED SCREW OF THE SAME BATCH, THE FOLLOWING ROOT CAUSE CAN BE IDENTIFIED: WHEN SPECIFIC COMBINATIONS OF GEOMETRICAL CONDITIONS MEET, THE PROTRUSION OF THE FIXING SCREW INSIDE THE LAG FLUSH PLATE IS LONGER THAN THE DEPTH OF THE FEMALE THREAD, CAUSING THE SCREW TO STOP ADVANCING DURING TIGHTENING AND BE CONSEQUENTLY STRESSED UNDER TORSION INSTEAD OF TRACTION. MODIFICATION OF THE IMPLANTS WILL BE MADE UNDER MR 038-23, TO PREVENT THE OCCURRENCE OF THIS ISSUE.

Description of Event or Problem · 0

DURING THE PRIMARY SPINE SURGERY, WHILE THE SURGEON WAS TIGHTENING THE ANTI BACK OUT LAG COVER PLATE, THE SURGEON THOUGHT HE HAD LOCKED THE PLATE. THE SURGEON THEN REMOVED THE PLATE HOLDER ALONG WITH THE DRIVER AND THE SURGEON THEN NOTICED THAT THE ANTIBACKOUT COVER PLATE WAS STILL ATTACHED TO THE PLATE HOLDER. THE SURGEON ALSO OBSERVED THAT THE TOP OF THE SCREW HAD BROKEN OFF FROM THE THREADED PORTION. THE SURGEON LEFT THE THREADED PORTION OF THE SCREW IN THE PLATE AND CONTINUED WITH THE CASE. NO POSTERIOR FIXATION PERFORMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1574641 MECTALIF ANTERIOR LAG PLATE FLUSH H12 MM SPINE ANTERIOR PLATE FLUSH OVD MEDACTA INTERNATIONAL SA 03.30.301 2220027 07630345732293

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other