FDA Adverse Event Malfunction Summary report: N

MECTALIF ANTERIOR LAG PLATE FLUSH H12 MM

MDR report key: 16511522 · Received March 9, 2023

Report

Report Number
3005180920-2023-00191
Event Type
Malfunction
Date Received
March 9, 2023
Date of Event
February 6, 2023
Report Date
April 6, 2023
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
OVD
UDI-DI
07630345732293
PMA / PMN Number
K221545
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

VISUAL INSPECTION PERFORMED BY R&D PROJECT MANAGER (ANALYSIS CONCLUDED ON 03-APRIL-2023). MECTALIF 03.30.10.2390 ALIF LAG TORQUE LIMITER STRAIGHT HANDLE 2.5NM - HL LOT: 2254942. THE TORQUE LIMITER HAS BEEN INSPECTED AND NO SIGN OF DAMAGE CAN BE FOUND. THE DEVICE HAS BEEN TESTED AND FOUND WITH THE CORRECT TOLERANCES OF 2.5 NM ± 10%. REF. 03.30.10.2300 ALIF T15 ANTI-BLACKOUT AND OFFSET SCREWDRIVER LOT: 2254973. BOTH SCREWDRIVERS HAVE BEEN INSPECTED AND NO SIGN OF DAMAGE CAN BE FOUND. THE PARTS ARE ACCORDING TO THE SPECIFICATIONS. REF. 03.30.301 MECTALIF ANT - LAG PLATE FLUSH H12 MM LOT: 2220027. THE FIXING SCREW OF THE COVER PLATE IS BROKEN. ANALYZING THE BREAKAGE SURFACE WITH A MICROSCOPE, SIGNS OF FRACTURE BY TORSION CAN BE SEEN. AFTER INTERNAL TESTS PERFORMED ON DEVICES OF THE SAME BATCH, THE FOLLOWING ROOT CAUSE CAN BE IDENTIFIED: WHEN SPECIFIC COMBINATIONS OF GEOMETRICAL CONDITIONS MEET, THE PROTRUSION OF THE FIXING SCREW INSIDE THE LAG FLUSH PLATE IS LONGER THAN THE DEPTH OF THE FEMALE THREAD, CAUSING THE SCREW TO STOP ADVANCING DURING TIGHTENING AND BE CONSEQUENTLY STRESSED UNDER TORSION INSTEAD OF TRACTION.

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 08-MAR-2023. LOT 2220027: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 21-OCT-2022. EXPIRATION DATE: 2027-08-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH ANOTHER REPORTED EVENT ON THE SAME LOT. OTHER DEVICE INVOLVED: MECTALIF 03.30.10.2390 ALIF LAG TORQUE LIMITER STRAIGHT HANDLE 2.5NM - HL LOT 2254942: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 08-NOV-2022. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, NO OTHER SIMILAR EVENTS HAVE BEEN REPORTED ON THE SAME LOT DURING THE PERIOD OF REVIEW. RECALL IS ON GOING.

Description of Event or Problem · 0

DURING THE PRIMARY SPINE SURGERY, WHEN THE SURGEON WAS ADDING THE ALIF ANTI-BACKOUT LAG PLATE, THE TORQUE LIMITER HANDLE DID NOT GIVE OFF A "CLICK" SOUND TO INDICATE WHEN TO STOP TORSIONAL PRESSURE/FORCE. AS A RESULT, THE SURGEON KEPT TRYING TO SCREW IN THE FACEPLATE AND THE ANTIBACKOUT SCREW BROKE IN HALF. THE DISTAL TIP OF THE ANTIBACKOUT SCREW COLD-FUSED INTO THE FACEPLATE. THE SURGEON LEFT THE 4 PRIMARY SCREWS AS THEY WERE AND COMPLETED THE CASE WITHOUT THE ANTIBACKOUT PLATE. THE SURGEON PERFORMED A POSTERIOR FIXATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1574640 MECTALIF ANTERIOR LAG PLATE FLUSH H12 MM SPINE ANTERIOR PLATE FLUSH OVD MEDACTA INTERNATIONAL SA 03.30.301 2220027 07630345732293

Patients

Seq Age Sex Outcome Treatment
1 Female Other