FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 2.0 TEST KIT 24T JAPAN

MDR report key: 16506594 · Received March 8, 2023

Report

Report Number
1221359-2023-00473
Event Type
Malfunction
Date Received
March 8, 2023
Date of Event
March 3, 2023
Report Date
May 24, 2023
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JP
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION RECEIVED: H4 - DEVICE MFG DATE AND D4 - EXPIRATION DATE THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 192-000J THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 192-000. TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT M217001 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 192-000J/ LOT: M217001, TEST BASE PART NUMBER 192-430/ LOT: M217001. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M217001 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, A POSSIBLE ASSIGNABLE ROOT CAUSE IS PATIENT SAMPLE INTERFERENCE. H3 OTHER TEXT : SINGLE USE, DEVICE DISCARDED.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 192-000J THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 192-000. THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. SINGLE USE, DEVICE DISCARDED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 2.0 ASSAY PERFORMED ON (B)(6) 2023 ON AN NASOPHARYNGEAL SAMPLE. ON THE SAME DAY PCR CONFIRMATION TESTING WAS PERFORMED TWICE USING PLATFORM: SHIMADZU'S AUTOAMP ON A NASOPHARYNGEAL SAMPLE AND GENERATED TWO NEGATIVE RESULTS RESPECTIVELY. THE CUSTOMER STATED THE PATIENT WAS ASYMPTOMATIC. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 2.0 ASSAY PERFORMED ON (B)(4) 2023 ON AN NASOPHARYNGEAL SAMPLE. ON THE SAME DAY PCR CONFIRMATION TESTING WAS PERFORMED TWICE USING PLATFORM: SHIMADZU'S AUTOAMP ON A NASOPHARYNGEAL SAMPLE AND GENERATED TWO NEGATIVE RESULTS RESPECTIVELY. THE CUSTOMER STATED THE PATIENT WAS ASYMPTOMATIC. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1727406 ID NOW COVID-19 2.0 TEST KIT 24T JAPAN REAGENTS, 2019-NOVEL CORONAVIRUS NUCLEIC ACID QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. M217001

Patients

Seq Age Sex Outcome Treatment
1 Unknown