FDA Adverse Event Malfunction Summary report: N

BD BBL¿ FLUID THIOGLYCOLLATE MEDIUM

MDR report key: 16504519 · Received March 8, 2023

Report

Report Number
1119779-2023-00244
Event Type
Malfunction
Date Received
March 8, 2023
Date of Event
February 28, 2023
Report Date
May 31, 2023
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
JSG
UDI-DI
30382902211960
PMA / PMN Number
PREAMENDMENT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: MATERIAL 221196 IS MANUFACTURED BY REHYDRATING THE MEDIA COMPONENTS WITH USP PURIFIED WATER, AND THOROUGHLY MIXING UNTIL A HOMOGENEOUS SOLUTION IS OBTAINED. THE MEDIA IS DISPENSED INTO BOTTLES; CAPS ARE APPLIED MANUALLY THEN TORQUED BY MACHINE PER A STANDARD OPERATING PROCEDURE (SOP). THE BOTTLES ARE THEN LABELED AND TERMINALLY AUTOCLAVED IN AN AIR OVER PRESSURE (AOP) AUTOCLAVE, PER MANUFACTURING INSTRUCTIONS, USING A VALIDATED CYCLE. POST AUTOCLAVING, BOTTLES ARE PACKAGED INTO FINAL SHIPPING CONFIGURATIONS. THE BATCH HISTORY RECORD REVIEW FOR BATCH 2312524 WAS SATISFACTORY AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. FORMULATION, FILLING, AND AUTOCLAVING PROCESSES WERE WITHIN SPECIFICATIONS. IN PROCESS CHECKS WERE PERFORMED AT THE DESIGNATED INTERVALS. THOSE CHECKS CONFIRMED THAT THE CAPS WERE TIGHTENED TO THE VALIDATED SPECIFICATIONS PER INTERNAL PROCEDURE. QC INSPECTION AND TESTING WAS SATISFACTORY AT TIME OF RELEASE. THE COMPLAINT HISTORY WAS REVIEWED, AND NO OTHER COMPLAINTS HAVE BEEN TAKEN ON THIS BATCH. RETENTION SAMPLES FROM BATCH 2312524 (10 TUBES) WERE AVAILABLE FOR INSPECTION. THE CERTIFICATE OF ANALYSIS STATES THE MEDIA APPEARANCE SHOULD BE LIGHT YELLOW WITH NO MORE THAN 1/3 OF FILL SHOWING ROSE PINK AT THE TOP OF THE MEDIUM AND MEDIUM IS CLEAR YELLOW WHEN HEATED. ALL 10/10 RETENTION SAMPLES MAINTAINED THE COLOR/APPEARANCE WITHIN THE SPECIFICATIONS DESCRIBED BY THE CERTIFICATE OF ANALYSIS. FOR FURTHER INVESTIGATION TWO RETENTION TUBES WERE INCUBATED. ONE TUBE WAS PLACED INTO 20-25-DEGREE CELSIUS, AND ONE RETENTION TUBE WAS PLACED INTO 33-37-DEGREES CELSIUS. AT THE END OF A SEVEN-DAY INCUBATION PERIOD NO MICROBIAL GROWTH OR PARTICLES WERE OBSERVED IN 2/2 RETENTION TUBES. TWO PHOTOS WERE RECEIVED TO ASSIST WITH THE INVESTIGATION: BOTH PHOTOS SHOW A PARTIAL TUBE FROM BATCH 2312524. THE CAP CANNOT BE SEEN IN BOTH PHOTOS. THE MEDIA APPEARS TO HAVE PARTICLES. NO RETURNS WERE RECEIVED TO ASSIST WITH THE INVESTIGATION. THE COMPLAINT CAN BE CONFIRMED BASED ON THE EVIDENCE PROVIDED BY THE PHOTOS RECEIVED. BD HAS IDENTIFIED A COMPLAINT TREND FOR HAZY MEDIA DUE TO THE PRESENCE OF NON-VIABLES FOR THIS PRODUCT. A CAPA (CORRECTIVE AND PREVENTATIVE ACTIONS) HAS BEEN INITIATED PER BD PROCEDURES. THE CAPA IS CURRENTLY IN THE INVESTIGATIONAL STAGE WHERE APPLICABLE CORRECTIVE ACTIONS ARE BEING IDENTIFIED. BD EXPECTS IMPROVEMENT IN THE OBSERVATION OF HAZY MEDIA DUE TO NON-VIABLES AS THE CAPA PROGRESSES. BD WILL CONTINUE TO TREND COMPLAINTS FOR CONTAMINATION. MATERIAL 221196 PRIOR TO USING IS TO BE BOILED FOR ONE MINUTE THE MEDIA SHOULD CLEAR TO A LIGHT YELLOW. THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D9: DEVICE AVAILABLE FOR EVALUATION:  YES; D9: RETURNED TO MANUFACTURER ON:  2023-04-17.

Additional Manufacturer Narrative · 0

COMMON DEVICE NAME: CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING THE BD BBL¿ FLUID THIOGLYCOLLATE MEDIUM THAT THERE WAS CONTAMINATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTS: AS PART OF OUR ACCEPTANCE PROCEDURES WE INCUBATE SEVERAL TUBES IMMEDIATELY ON RECEPTION AS A NEGATIVE CONTROL BEFORE GROWTH PROMOTION IS COMPLETED. ONE OF THE FTM TUBES THAT WE INCUBATED CAME BACK POSITIVE DESPITE BEING A NEGATIVE CONTROL AND WE FOUND PARTICULATES FLOATING IN THE MEDIA OF 2 OTHER TUBES. LOT 2312524.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING THE BD BBL¿ FLUID THIOGLYCOLLATE MEDIUM THAT THERE WAS CONTAMINATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTS: AS PART OF OUR ACCEPTANCE PROCEDURES WE INCUBATE SEVERAL TUBES IMMEDIATELY ON RECEPTION AS A NEGATIVE CONTROL BEFORE GROWTH PROMOTION IS COMPLETED. ONE OF THE FTM TUBES THAT WE INCUBATED CAME BACK POSITIVE DESPITE BEING A NEGATIVE CONTROL AND WE FOUND PARTICULATES FLOATING IN THE MEDIA OF 2 OTHER TUBES. LOT 2312524.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1706393 BD BBL¿ FLUID THIOGLYCOLLATE MEDIUM SEE H.10 JSG BECTON, DICKINSON & CO. (SPARKS) 221196 2312524 30382902211960

Patients

Seq Age Sex Outcome Treatment
1 Unknown