FDA Adverse Event Malfunction Summary report: N

BD ULTRA-FINE¿ II SHORT NEEDLE INSULIN SYRINGE 0.3ML 0.30MM X 8MM

MDR report key: 16502854 · Received March 8, 2023

Report

Report Number
1920898-2023-00098
Event Type
Malfunction
Date Received
March 8, 2023
Date of Event
February 9, 2023
Report Date
March 17, 2023
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY NO SAMPLES WERE RETURNED THEREFORE THE INVESTIGATION WAS PERFORMED BASED ON THE PHOTOS PROVIDED. THE CUSTOMER RETURNED THREE PHOTOS OF THE 30GX8MM POLYBAGS FROM LOT 2173943. THE CUSTOMER REPORTED POLYBAG TORN AND LACK OF INFORMATION. THE PHOTOS WERE EXAMINED, AND IT WAS OBSERVED THAT THE POLYBAG IS TORN DOWN THE BACK OF THE POUCH. ANOTHER PHOTO REVEALS THE LOT NUMBER, EXPIRATION DATE AND MANUFACTURER DATE MISSING FROM THE POLYBAG. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 2173943. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. BASED ON THE PHOTOS RECEIVED, EMBECTA WAS ABLE TO CONFIRM THE CUSTOMER¿S INDICATED FAILURE MISSING LABEL CONTENT (LOT NUMBER, MANUFACTURER DATE AND EXPIRATION DATE MISSING) A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD ULTRA-FINE¿ II SHORT NEEDLE INSULIN SYRINGE 0.3ML 0.30MM (30G) X 8MM (5/16") U-100 HAD MISSING LABEL INFORMATION. IT WAS REPORTED THAT 400 WERE AFFECTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MISSING INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD ULTRA-FINE¿ II SHORT NEEDLE INSULIN SYRINGE 0.3ML 0.30MM (30G) X 8MM (5/16") U-100 HAD MISSING LABEL INFORMATION. IT WAS REPORTED THAT 400 WERE AFFECTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MISSING INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
136282 BD ULTRA-FINE¿ II SHORT NEEDLE INSULIN SYRINGE 0.3ML 0.30MM X 8MM PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE 2173943

Patients

Seq Age Sex Outcome Treatment
1 Unknown