FDA Adverse Event Malfunction Summary report: N

ORTHO SUMMIT SAMPLE HANDLING SYSTEM

MDR report key: 1650234 · Received April 5, 2010

Report

Report Number
2250051-2010-00086
Event Type
Malfunction
Date Received
April 5, 2010
Date of Event
April 1, 2010
Report Date
April 5, 2010
Manufacturer
HAMILTON BONADUZ AG
Product Code
JTC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT HAS BEEN INVESTIGATED. THE FIELD SERVICE ENGINEER (FSE) INSPECTED THE INSTRUMENT AND FOUND A STUCK EJECT PIN IN PLUNGER CLAMP #8. THE FSE CLEANED THE PLUNGER CLAMP. THE FSE PERFORMED SPLLD CHECKING PROCEDURE AND TESTED SUMMIT WITH OAS USER SOFTWARE. THE INSTRUMENT WAS TESTED WITHOUT PROBLEM AND WAS RETURNED TO EXPECTED OPERATION.

Description of Event or Problem · 1

THE ORTHO SUMMIT SAMPLE HANDLING SYSTEM INSTRUMENT REPORTEDLY DID NOT PIPETTE SAMPLE AND REAGENT AND DID NOT GENERATE AN ERROR MESSAGE.NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT.THE REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ORTHO SUMMIT SAMPLE HANDLING SYSTEM SAMPLE PIPETTER JTC HAMILTON BONADUZ AG 936400

Patients

Seq Age Sex Outcome Treatment
1