ORTHO SUMMIT SAMPLE HANDLING SYSTEM
MDR report key: 1650136
·
Received April 5, 2010
Report
- Report Number
- 2250051-2010-00085
- Event Type
- Malfunction
- Date Received
- April 5, 2010
- Date of Event
- March 30, 2010
- Report Date
- April 5, 2010
- Manufacturer
- HAMILTON BONADUZ AG
- Product Code
- JTC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT HAS BEEN INVESTIGATED. THE FIELD SERVICE ENGINEER (FSE) INSPECTED THE INSTRUMENT AND FOUND PLUNGER CLAMP EJECT PIN #12 STUCK. THE FSE CLEANED THE PLUNGER CLAMP AND EXERCISED THE EJECT SPRING. THE FSE PERFORMED SPLLD CHECKING PROCEDURE AND TESTED SUMMIT WITH OAS USER SOFTWARE. THE INSTRUMENT WAS TESTED WITHOUT PROBLEM AND WAS RETURNED TO EXPECTED OPERATION.
Description of Event or Problem · 1
THE ORTHO SUMMIT SAMPLE HANDLING SYSTEM INSTRUMENT REPORTEDLY DID NOT PIPETTE REAGENT AND DID NOT GENERATE AN ERROR MESSAGE.NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT.THE REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (B)(4).