FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK¿ TIP SYRINGE

MDR report key: 16500542 · Received March 8, 2023

Report

Report Number
1911916-2023-00124
Event Type
Malfunction
Date Received
March 8, 2023
Date of Event
February 2, 2023
Report Date
March 30, 2023
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMF
UDI-DI
00382903096282
PMA / PMN Number
K941562
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES, D10: RETURNED TO MANUFACTURER ON: 23-MAR-2023. INVESTIGATION SUMMARY: ONE SAMPLE WAS PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, IT WAS OBSERVED THAT THE TIP OF THE PLUNGER ROD WAS BROKEN INSIDE THE STOPPER. POTENTIAL ROOT CAUSE FOR THE BROKEN PLUNGER ROD DEFECT IS ASSOCIATED WITH THE ASSEMBLY PROCESS. THESE DEFECTS ARE OCCURRING BELOW AN EXPECTED RATE SO NO CORRECTIVE ACTIONS WILL BE MADE AT THIS TIME. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 2213779. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE WITH BD LUER-LOK¿ TIP SYRINGE THE PLUNGER BROKE. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: DURING INJECTION OF METHOTREXATE INTO A 100 ML FREEFLEX NACL BAG, THE PLUNGER OF THE 1 ML SYRINGE BROKE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE WITH BD LUER-LOK¿ TIP SYRINGE THE PLUNGER BROKE. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: DURING INJECTION OF METHOTREXATE INTO A 100 ML FREEFLEX NACL BAG, THE PLUNGER OF THE 1 ML SYRINGE BROKE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
122212 BD LUER-LOK¿ TIP SYRINGE PISTON SYRINGE FMF BD MEDICAL (BD WEST) MEDICAL SURGICAL 2213779 00382903096282

Patients

Seq Age Sex Outcome Treatment
1 Unknown