FDA Adverse Event Malfunction Summary report: N

VCL+ UD 27IN 5-0 S/A RB-1

MDR report key: 16500507 · Received March 7, 2023

Report

Report Number
2210968-2023-01513
Event Type
Malfunction
Date Received
March 7, 2023
Date of Event
February 14, 2023
Report Date
March 27, 2023
Manufacturer
ETHICON INC.
Product Code
GAM
UDI-DI
10705031051645
PMA / PMN Number
K132580
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). COMPONENT CODE: G07002 - DEVICE NOT RETURNED. THIS IS A COMBINATION PRODUCT, AND THE EVENT HAS BEEN REVIEWED FOR BOTH THE SUTURE AND THE TRICLOSAN. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803, PART 4 SUBPART B. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. ATTEMPTS HAVE BEEN MADE TO RETRIEVE THE DEVICE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT, AND NO NON-CONFORMANCES WERE IDENTIFIED. TRADE NAME - IRGACARE®, ACTIVE INGREDIENT(S) ¿ TRICLOSAN, DOSAGE FORM ¿ SUTURE/SOLID/PARENTERAL AND STRENGTH ¿ = 275 G/M.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). DATE SENT TO THE FDA: 3/27/2023. THIS IS A COMBINATION PRODUCT, AND THE EVENT HAS BEEN REVIEWED FOR BOTH THE SUTURE AND THE TRICLOSAN. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803, PART 4 SUBPART B. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. H6 COMPONENT CODE: G07002 NO DEVICE PROBLEM FOUND. H3 INVESTIGATIONAL SUMMARY: THE PRODUCT WAS RETURNED TO ETHICON FOR EVALUATION. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE CONDUCTED ON THE RETURNED DEVICE. THE RETURNED PRODUCT DETERMINED THAT IT WAS RECEIVED ONE NEEDLE SUTURE IN TWO SECTIONS THAT PERTAIN TO PRODUCT CODE VCP213. OVERALL REVIEW OF THE RETURNED SAMPLE SHOWS THAT THE SWAGE AND ATTACHMENT AREA OF THE NEEDLE WAS NOTED TO BE AS EXPECTED. THE NEEDLE IS STILL ATTACHED TO A SUTURE PIECE. IN ADDITION, THE END OF THE SUTURE PIECE WAS OBSERVED DAMAGE. DUE TO RETURNED CONDITION OF THE SAMPLE, NO FUNCTIONAL TEST COULD BE PERFORMED. AS PART OF OUR QUALITY PROCESS, THE MANUFACTURING RECORDS OF THIS LOT-SERIAL NUMBER WERE REVIEWED, AND THE MANUFACTURING STANDARDS WERE MET PRIOR TO THE RELEASE OF THIS BATCH. AS PART OF THE ETHICON QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS IS CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST-MARKET SURVEILLANCE.

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT AN EMERGENCY DEBRIDEMENT PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
136866 VCL+ UD 27IN 5-0 S/A RB-1 SUTURE, ABSORBABLE, SYNTHETIC GAM ETHICON INC. VCP213H SEMRQR 10705031051645

Patients

Seq Age Sex Outcome Treatment
1 Unknown