FDA Adverse Event Malfunction Summary report: N

ARCHITECT TOTAL B-HCG REAGENT KIT

MDR report key: 16500113 · Received March 7, 2023

Report

Report Number
3005094123-2023-00065
Event Type
Malfunction
Date Received
March 7, 2023
Date of Event
February 10, 2023
Report Date
March 14, 2023
Manufacturer
A.I.D.D LONGFORD
Product Code
DHA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AFTER FURTHER EVALUATION, THE SUSPECT MEDICAL DEVICE WAS CHANGED FROM ARCHITECT TOTAL B-HCG, LIST 07K78-77, A.I.D.D LONGFORD, LONGFORD, IRELAND TO THE ARCHITECT I2000SR PROCESSING MODULE, LIST 03M74-02, ABBOTT LABORATORIES, IRVING, TEXAS, USA. MDR NUMBER 3016438761-2023-00140-00 HAS BEEN SUBMITTED AND ALL FURTHER INFORMATION WILL BE DOCUMENTED UNDER THAT MDR NUMBER.

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSE POSITIVE ARCHITECT TOTAL B-HCG RESULT ON A PATIENT WHO WAS DIAGNOSED WITH PARTIAL HYDATIDIFORM MOLE AND HAD UNDERGONE SURGICAL RESECTION. THE RESULT WAS QUESTIONED BY THE PHYSICIAN AS IT WAS NOT CONSISTENT WITH THE PATIENT¿S CLINICAL DIAGNOSIS. A NEW SAMPLE WAS TESTED, AND THE RESULT WAS NEGATIVE. THE ORIGINAL SAMPLE WAS ALSO RETESTED, AND THE RESULT WAS NEGATIVE. THE FOLLOWING DATA WAS PROVIDED: ORIGINAL SAMPLE: INITIAL RESULT = 153.72 MIU/ML (POSITIVE); REPEAT RESULT = < 1.2 MIU/ML (NEGATIVE). NEW SAMPLE: RESULT = < 1.2 MIU/ML (NEGATIVE). COLLOIDAL GOLD TEST = NEGATIVE. ADDITIONAL INFORMATION RECEIVED ON 03MAR2023. IT WAS NOTED THAT THE FILTER TO THE CUSTOMER¿S WATER SYSTEM HAD NOT BEEN REPLACED FOR A LONG TIME. IT IS SUSPECTED THAT THE CUSTOMER'S WATER MAY HAVE CONTRIBUTED TO THE DISCREPANT RESULT. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1776565 ARCHITECT TOTAL B-HCG REAGENT KIT SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN DHA A.I.D.D LONGFORD 42330UD02

Patients

Seq Age Sex Outcome Treatment
1 Unknown ARC I2000SR INST, 03M74-02, ISR64377| ARC I2000SR INST, 03M74-02, ISR64377