RESONATE X4 CRT-D
Report
- Report Number
- 2124215-2023-10797
- Event Type
- Malfunction
- Date Received
- March 7, 2023
- Date of Event
- February 2, 2022
- Report Date
- April 14, 2023
- Manufacturer
- BOSTON SCIENTIFIC CORPORATION
- Product Code
- NIK
- PMA / PMN Number
- P010012/S436
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
IT WAS REPORTED THAT OVER THE COURSE OF FIVE DAYS, THIS PATIENT RECEIVED THREE INAPPROPRIATE SHOCKS DUE TO DOUBLE-COUNTING OVER-SENSING. BOSTON SCIENTIFIC TECHNICAL SERVICES WAS CONTACTED FOR A DEVICE DATA REVIEW. IT WAS DISCUSSED THAT THE PATIENT'S RIGHT VENTRICULAR (RV) R-WAVE MEASUREMENTS WERE QUITE LOW, WHICH CONTRIBUTED TO THE DOUBLE-COUNTING. THE PHYSICIAN HAD STATED THAT THE PATIENT RECENTLY HAD AN ABLATION PROCEDURE, WHICH WAS STATED TO BE THE CAUSE OF THE LOW R-WAVE AMPLITUDE MEASUREMENTS. THE PHYSICIAN ELECTED TO ALTER THE PATIENT'S MEDICATION AND PROGRAMMED THERAPY SETTINGS TO RESOLVE THE EVENT. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. THIS DEVICE REMAINS IMPLANTED AND IN-SERVICE.
IT WAS REPORTED THAT OVER THE COURSE OF FIVE DAYS, THIS PATIENT RECEIVED THREE INAPPROPRIATE SHOCKS DUE TO DOUBLE-COUNTING OVER-SENSING. BOSTON SCIENTIFIC TECHNICAL SERVICES WAS CONTACTED FOR A DEVICE DATA REVIEW. IT WAS DISCUSSED THAT THE PATIENT'S RIGHT VENTRICULAR (RV) R-WAVE MEASUREMENTS WERE QUITE LOW, WHICH CONTRIBUTED TO THE DOUBLE-COUNTING. THE PHYSICIAN HAD STATED THAT THE PATIENT RECENTLY HAD AN ABLATION PROCEDURE, WHICH WAS STATED TO BE THE CAUSE OF THE LOW R-WAVE AMPLITUDE MEASUREMENTS. THE PHYSICIAN ELECTED TO ALTER THE PATIENT'S MEDICATION AND PROGRAMMED THERAPY SETTINGS TO RESOLVE THE EVENT. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. THIS DEVICE REMAINS IMPLANTED AND IN-SERVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 125596 | RESONATE X4 CRT-D | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D | NIK | BOSTON SCIENTIFIC CORPORATION | G447 | 278834 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |