FDA Adverse Event Malfunction Summary report: N

BD BBL¿ SALMONELLA SHIGELLA AGAR PLATE 20PK

MDR report key: 16499208 · Received March 7, 2023

Report

Report Number
1119779-2023-00238
Event Type
Malfunction
Date Received
March 7, 2023
Date of Event
February 22, 2023
Report Date
June 12, 2024
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
JSI
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY MATERIAL #: [251181]LOT/BATCH #: #2325441 BD RECEIVED PHOTOS FOR EVALUATION. BD WAS ABLE TO DETERMINE THE ISSUE WAS PRECIPITATION. THIS COMPLAINT IS ABLE TO BE CONFIRMED. THE COMPLAINT HISTORY HAS BEEN REVIEWED AND NO TREND WAS IDENTIFIED. THE BATCH HISTORY REVIEW DID NOT INDICATE ANY DISCREPANCIES. ALL RELEASE TESTING WAS SATISFACTORY, AND NO DEVIATIONS WERE OBSERVED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD BBL¿ SALMONELLA SHIGELLA AGAR PLATE 20PK MOLD WAS FOUND IN THE MEDIA. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ACCORDING TO THE CUSTOMER'S REPORT, BACTERIAL RESEMBLING MOLD WAS FOUND ON THE MEDIA.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD BBL¿ SALMONELLA SHIGELLA AGAR PLATE 20PK MOLD WAS FOUND IN THE MEDIA. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ACCORDING TO THE CUSTOMER'S REPORT, BACTERIAL RESEMBLING MOLD WAS FOUND ON THE MEDIA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
86504 BD BBL¿ SALMONELLA SHIGELLA AGAR PLATE 20PK CULTURE MEDIA, SELECTIVE AND DIFFERENTIAL JSI BECTON, DICKINSON & CO. (SPARKS) 2325441

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown