ELECSYS ANTI-CCP
Report
- Report Number
- 1823260-2023-00712
- Event Type
- Malfunction
- Date Received
- March 7, 2023
- Date of Event
- January 30, 2023
- Report Date
- May 9, 2023
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- NHX
- UDI-DI
- 04015630940370
- PMA / PMN Number
- K081338
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ONE PATIENT SAMPLE WAS PROVIDED FOR INVESTIGATION AND THE CUSTOMER'S RESULTS COULD BE REPRODUCED. NA.
THE PATIENT IS HEALTHY AND PARTICIPATES IN A LOT OF SPORTS. THE PATIENT HAD NEGATIVE RESULTS FOR CRP AND RHEUMATOID FACTOR. THE LAST CALIBRATION PERFORMED ON 14-DEC-2022 WAS WITHIN EXPECTATIONS. QUALITY CONTROLS RECOVERED WITHIN SPECIFIED RANGES. FURTHER INVESTIGATIONS OF THE PATIENT SAMPLE DETERMINED THAT IT CONTAINS A POTENTIAL INTERFERING FACTOR AGAINST A COMPONENT OF ONE LOT OF ANTI-CCP REAGENT (LOT 637426), RESULTING IN A FALSE LOW ANTI-CCP VALUE. FURTHER CLARIFICATION OF THE OBSERVED DISCREPANCY WAS NOT POSSIBLE WITH AVAILABLE METHODS AND THE CURRENT STATE OF THE ART.
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR TWO SAMPLES COLLECTED FROM ONE PATIENT AND TESTED WITH ELECSYS ANTI-CCP ON A COBAS 6000 E 601 MODULE (SERIAL NUMBER (B)(4) AND A COBAS E 801 ANALYTICAL UNIT (SERIAL NUMBER (B)(4). THE SAMPLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY TO THE PATIENT. WHEN THE FIRST SAMPLE WAS TESTED ON THE E 801 ANALYZER, IT RESULTED IN AN ANTI-CCP VALUE OF 41.1 U/ML ON (B)(6) 2023. THIS SAMPLE WAS REPEATED ON THE E 601 ANALYZER ON (B)(6) 2023, RESULTING IN A VALUE OF < 7.00 U/ML WITH A DATA FLAG. THE SECOND SAMPLE HAD SIMILAR RESULTS. ON (B)(6) 2023, THE SECOND SAMPLE RESULTED IN AN ANTI-CCP VALUE OF 38.0 U/ML WHEN TESTED ON THE E 801 ANALYZER. THIS SAMPLE RESULTED IN A VALUE OF < 7.00 U/ML WHEN TESTED ON THE E 601 ANALYZER. ANTI-CCP REAGENT LOT 66022801 (EXPIRATION DATE = 30-SEP-2023) WAS USED ON THE E 801 ANALYZER. ANTI-CCP REAGENT LOT 637426 WAS USED ON THE E 601 ANALYZER. THIS REAGENT EXPIRATION DATE WAS REQUESTED, BUT NOT PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 716583 | ELECSYS ANTI-CCP | ANTIBODIES, ANTI-CYCLIC CITRULLINATED PEPTIDE (CCP) | NHX | ROCHE DIAGNOSTICS | NA | 66022801, 637426 | 04015630940370 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 22 YR | Female |