FDA Adverse Event Malfunction Summary report: N

ELECSYS ANTI-CCP

MDR report key: 16498612 · Received March 7, 2023

Report

Report Number
1823260-2023-00712
Event Type
Malfunction
Date Received
March 7, 2023
Date of Event
January 30, 2023
Report Date
May 9, 2023
Manufacturer
ROCHE DIAGNOSTICS
Product Code
NHX
UDI-DI
04015630940370
PMA / PMN Number
K081338
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ONE PATIENT SAMPLE WAS PROVIDED FOR INVESTIGATION AND THE CUSTOMER'S RESULTS COULD BE REPRODUCED. NA.

Additional Manufacturer Narrative · 0

THE PATIENT IS HEALTHY AND PARTICIPATES IN A LOT OF SPORTS. THE PATIENT HAD NEGATIVE RESULTS FOR CRP AND RHEUMATOID FACTOR. THE LAST CALIBRATION PERFORMED ON 14-DEC-2022 WAS WITHIN EXPECTATIONS. QUALITY CONTROLS RECOVERED WITHIN SPECIFIED RANGES. FURTHER INVESTIGATIONS OF THE PATIENT SAMPLE DETERMINED THAT IT CONTAINS A POTENTIAL INTERFERING FACTOR AGAINST A COMPONENT OF ONE LOT OF ANTI-CCP REAGENT (LOT 637426), RESULTING IN A FALSE LOW ANTI-CCP VALUE. FURTHER CLARIFICATION OF THE OBSERVED DISCREPANCY WAS NOT POSSIBLE WITH AVAILABLE METHODS AND THE CURRENT STATE OF THE ART.

Description of Event or Problem · 0

THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR TWO SAMPLES COLLECTED FROM ONE PATIENT AND TESTED WITH ELECSYS ANTI-CCP ON A COBAS 6000 E 601 MODULE (SERIAL NUMBER (B)(4) AND A COBAS E 801 ANALYTICAL UNIT (SERIAL NUMBER (B)(4). THE SAMPLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY TO THE PATIENT. WHEN THE FIRST SAMPLE WAS TESTED ON THE E 801 ANALYZER, IT RESULTED IN AN ANTI-CCP VALUE OF 41.1 U/ML ON (B)(6) 2023. THIS SAMPLE WAS REPEATED ON THE E 601 ANALYZER ON (B)(6) 2023, RESULTING IN A VALUE OF < 7.00 U/ML WITH A DATA FLAG. THE SECOND SAMPLE HAD SIMILAR RESULTS. ON (B)(6) 2023, THE SECOND SAMPLE RESULTED IN AN ANTI-CCP VALUE OF 38.0 U/ML WHEN TESTED ON THE E 801 ANALYZER. THIS SAMPLE RESULTED IN A VALUE OF < 7.00 U/ML WHEN TESTED ON THE E 601 ANALYZER. ANTI-CCP REAGENT LOT 66022801 (EXPIRATION DATE = 30-SEP-2023) WAS USED ON THE E 801 ANALYZER. ANTI-CCP REAGENT LOT 637426 WAS USED ON THE E 601 ANALYZER. THIS REAGENT EXPIRATION DATE WAS REQUESTED, BUT NOT PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
716583 ELECSYS ANTI-CCP ANTIBODIES, ANTI-CYCLIC CITRULLINATED PEPTIDE (CCP) NHX ROCHE DIAGNOSTICS NA 66022801, 637426 04015630940370

Patients

Seq Age Sex Outcome Treatment
1 22 YR Female