RAYSTATION
Report
- Report Number
- 3010034862-2023-00023
- Event Type
- Malfunction
- Date Received
- March 7, 2023
- Date of Event
- December 5, 2019
- Report Date
- March 31, 2025
- Manufacturer
- RAYSEARCH LABORATORIES AB (PUBL)
- Product Code
- MUJ
- Product Problem
- Yes
- Report Source
- Distributor report
- Reporter Location
- SW
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
FACTORS THAT MAY HAVE CAUSED OR CONTRIBUTED TO THE ADVERSE EVENT AND THE ACTUAL OR POTENTIAL HEALTH HAZARD: A SOFTWARE IMPLEMENTATION ERROR HAS BEEN IDENTIFIED. THE IMPLEMENTATION DOES NOT DISTINGUISH CORRECTLY BETWEEN SSD DEFINED AS SOURCE-TO-SKIN AND SSD DEFINED AS SOURCE-TO-SURFACE. IMMEDIATE AND/OR LONG-RANGE HEALTH CONSEQUENCES OF THE ACTUAL OR POTENTIAL HEALTH HAZARD: IN THE EVENT THAT THE GETSSD SCRIPTING FUNCTION IS INCLUDED IN A SCRIPT THAT IS SOMEHOW USED TO INFLUENCE CLINICAL DECISIONS, SSD INPUT DATA COULD BE MISLEADING.
NO PATIENTS WERE AFFECTED. INTERNALLY FOUND BUG IN THE SCRIPTABLE METHOD BEAM.GETSSD(). THE SCRIPT METHOD BEAM.GETSSD() CAN RETURN SOURCE-TO-SKIN DISTANCE INSTEAD OF THE EXPECTED SOURCE-TO-SURFACE DISTANCE, DEPENDING ON WHEN IT IS RUN. THE SAME ISSUE AFFECTS THE SSD DISPLAY IN DRR VIEWS IN THE REPORT AND IN EXPORTED DRRS. IN ADDITION, IN RAYSTATION 4.0, 4.5 AND 4.3 INVERSEARC, THE SSD IN THE REPORT DRR MAY SHOW OTHER VALUES THAN THE SOURCE-TO-SKIN OR SOURCE-TO-SURFACE DISTANCE. THE MAGNITUDE OF THE ERROR WILL DEPEND ON THE EXTENSION OF ANY BOLUS, FIXATION OR SUPPORT ROI IN THE BEAM CENTER PATH. THE ISSUE AFFECTS ELECTRON, PHOTON AND BNCT BEAMS ONLY, SINCE BEAM.GETSSD() CANNOT BE USED FOR ION BEAMS AND SSD IS NOT DISPLAYED IN THE DRR FOR ION BEAMS. ONLY SCRIPTING, EXPORTED DRRS AND DRR IMAGES IN THE PLAN REPORT ARE AFFECTED. SSD IN THE PLAN REPORT IS CORRECT IN THE BEAM DATA SECTION, WHERE BOTH THE SOURCE-TO-SKIN AND SOURCE-TO-SURFACE VALUES ARE DISPLAYED. SSD IS ALSO CORRECT IN THE RAYSTATION GRAPHICAL USER INTERFACE (GUI). FOR SETUP BEAMS, THE DRRS SHOW CORRECT VALUE, BUT THE SCRIPT METHOD IS AFFECTED BY THE ISSUE.
FOLLOW-UP OF REPORT RSL: MDR 3010034862-2023-00023 52971 RS SSD INCONSISTENCY IN SCRIPTING METHOD GETSSD AND DRR IN REPORT. NO PATIENTS WERE AFFECTED. INTERNALLY FOUND BUG IN THE SCRIPTABLE METHOD BEAM.GETSSD(). THE SCRIPT METHOD BEAM.GETSSD() CAN RETURN SOURCE-TO-SKIN DISTANCE INSTEAD OF THE EXPECTED SOURCE-TO-SURFACE DISTANCE, DEPENDING ON WHEN IT IS RUN. THE SAME ISSUE AFFECTS THE SSD DISPLAY IN DRR VIEWS IN THE REPORT AND IN EXPORTED DRRS. IN ADDITION, IN RAYSTATION 4.0, 4.5 AND 4.3 INVERSEARC, THE SSD IN THE REPORT DRR MAY SHOW OTHER VALUES THAN THE SOURCE-TO-SKIN OR SOURCE-TO-SURFACE DISTANCE. THE MAGNITUDE OF THE ERROR WILL DEPEND ON THE EXTENSION OF ANY BOLUS, FIXATION OR SUPPORT ROI IN THE BEAM CENTER PATH. THE ISSUE AFFECTS ELECTRON, PHOTON AND BNCT BEAMS ONLY, SINCE BEAM.GETSSD() CANNOT BE USED FOR ION BEAMS AND SSD IS NOT DISPLAYED IN THE DRR FOR ION BEAMS. ONLY SCRIPTING, EXPORTED DRRS AND DRR IMAGES IN THE PLAN REPORT ARE AFFECTED. SSD IN THE PLAN REPORT IS CORRECT IN THE BEAM DATA SECTION, WHERE BOTH THE SOURCE-TO-SKIN AND SOURCE-TO-SURFACE VALUES ARE DISPLAYED. SSD IS ALSO CORRECT IN THE RAYSTATION GRAPHICAL USER INTERFACE (GUI). FOR SETUP BEAMS, THE DRRS SHOW CORRECT VALUE, BUT THE SCRIPT METHOD IS AFFECTED BY THE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 63088 | RAYSTATION | RADIATION THERAPY TREATMENT PLANNING SYSTEM | MUJ | RAYSEARCH LABORATORIES AB (PUBL) | RAYSTATION 4-9A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |