BD INTIMA-II IV CATHETER
Report
- Report Number
- 3014704491-2023-00031
- Event Type
- Malfunction
- Date Received
- March 7, 2023
- Date of Event
- February 6, 2023
- Report Date
- February 16, 2023
- Manufacturer
- BD (SUZHOU)
- Product Code
- FOZ
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INITIAL REPORTER PHONE #: (B)(6). INVESTIGATION SUMMARY: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 0140311. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT. BD WILL CONTINUE TO MONITOR THIS ISSUE AND ENCOURAGES YOU TO SUBMIT YOUR SAMPLE FOR REVIEW.
IT WAS REPORTED THAT THE BD INTIMA-II IV CATHETER EXPERIENCED LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ON (B)(6) 2023, THE PATIENT UNDERWENT LEFT BREAST LUMPECTOMY UNDER LOCAL ANESTHESIA. DURING THE PREOPERATIVE REHYDRATION PROCESS WITH AN INTRAVENOUS INDWELLING NEEDLE, IT WAS FOUND THAT THE HEPARIN CAP OF THE INTRAVENOUS INDWELLING NEEDLE WAS LEAKING. IMMEDIATELY AFTER REPLACING THE NEW HEPARIN CAP, THE FLUID CAN BE REPLENISHED NORMALLY WITHOUT CAUSING ADVERSE EFFECTS ON THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 86234 | BD INTIMA-II IV CATHETER | INTRAVASCULAR CATHETER | FOZ | BD (SUZHOU) | 0140311 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |