FDA Adverse Event Malfunction Summary report: N

VERSO AIRWAY ACCESS ADAPTER

MDR report key: 1649635 · Received March 19, 2010

Report

Report Number
1649635
Event Type
Malfunction
Date Received
March 19, 2010
Date of Event
March 10, 2010
Report Date
March 19, 2010
Manufacturer
CAREFUSION AIRLIFE
Product Code
BZA
Product Problem
Yes
Report Source
User Facility report
Reporter Location
MD, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

PT IN ICU ON VENTILATOR. WHILE PATIENT WAS BEING TURNED THE NURSE HEARD AIR LEAK. THE VERSO ADAPTER HAD SNAPPED IN TWO AND PART OF IT REMAINED IN THE PATIENT. THE PT. BECAME DISCONNECTED FROM THE VENTILATOR. ONE NURSE HELD ON TO THE ADAPTER AND KEEPING INLINE SUCTION TOGETHER WHILE ANOTHER WAS USING AMBU BAG TO VENTILATE PATIENT. PIECE WAS REMOVED FROM PATIENT'S THROAT WITH FORCEPS WITHOUT INJURY. INLINE SUCTION WAS REPLACED.====================== HEALTH PROFESSIONAL'S IMPRESSION======================NO ADVERSE EVENT. INLINE ADAPTER SNAPPED, OPENING THE CLOSED SYSTEM====================== MANUFACTURER RESPONSE FOR ADAPTER FOR VENTILATOR INLINE SUCTION, VERSO AIRWAY ACCESS ADAPTER======================"WE HAVE HAD ONLY ONE VERSO SNAP BEFORE..."IT WAS TESTED AND APPEARED FINE. MAKING SEVERAL MODIFICATIONS TO DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VERSO AIRWAY ACCESS ADAPTER ADAPTER FOR VENTILATOR INLINE SUCTION BZA CAREFUSION AIRLIFE CE0086 09060014

Patients

Seq Age Sex Outcome Treatment
1 45 YR OTHER