FDA Adverse Event Malfunction Summary report: N

KONG-TL VBR SYSTEM, SCREWDRIVER FOR LOCKING SCREW

MDR report key: 16495022 · Received March 7, 2023

Report

Report Number
3006493760-2023-00001
Event Type
Malfunction
Date Received
March 7, 2023
Date of Event
January 30, 2023
Report Date
March 7, 2023
Manufacturer
ICOTEC AG
Product Code
MQP
UDI-DI
07640164847849
PMA / PMN Number
K200235
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 0

DURING A CORPECTOMY AN EXPANDABLE ICOTEC VBR IMPLANT WAS IMPLANTED. THE SET HEIGHT OF THE VBR IS FIXED WITH A LOCKING SCREW. WHEN TIGHTENING THE LOCKING SCREW, IT WAS NOTICED THAT THE SCREWDRIVER TIP HAD BROKEN OFF. THE BROKEN OFF FRAGMENT HAS STUCK IN THE HEAD OF THE LOCKING SCREW. THE VBR WAS LEFT IMPLANTED AND THE SURGERY WAS COMPLETED WITHOUT FURTHER COMPLICATIONS. THE BROKEN OFF FRAGMENT REMAINED IN THE SCREW HEAD AND THEREFORE IN THE PATIENT'S BODY. THE SUITABILITY OF THE MATERIAL OF THE FRAGMENT (STAINLESS STEEL 1.4112) HAS NOT BEEN EVALUATED FOR LONG-TERM TISSUE/BONE CONTACT. UNDESIRABLE LOCAL OR SYSTEMIC EFFECTS CANNOT BE EXCLUDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
140542 KONG-TL VBR SYSTEM, SCREWDRIVER FOR LOCKING SCREW Spinal vertebral body replacement device MQP ICOTEC AG 44-050 20/02 07640164847849

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other