RAYSTATION
Report
- Report Number
- 3007774465-2023-00015
- Event Type
- Malfunction
- Date Received
- March 6, 2023
- Date of Event
- October 13, 2020
- Report Date
- March 3, 2025
- Manufacturer
- RAYSEARCH LABORATORIES AB (PUBL)
- Product Code
- MUJ
- PMA / PMN Number
- K200569
- Removal / Correction Number
- RES 86832
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
A SOFTWARE IMPLEMENTATION ERROR HAS BEEN IDENTIFIED. IF THE INCORRECT ANGLES ARE USED FOR INITIAL SETUP ONLY AND THE INCORRECT SETUP IS THEN DETECTED USING IMAGING, THERE COULD BE UNNECESSARY IMAGING DOSE IF REPEAT IMAGES ARE REQUIRED AFTER CORRECTION. IF THE INCORRECT PITCH/ROLL VALUES ARE NOT DETECTED AND USED FOR TREATMENT, THIS COULD LEAD TO LOCAL OVER-DOSE IN A RISK ORGAN OR LOCAL UNDER-DOSAGE TO THE TARGET. SEVERITY WOULD DEPEND ON THE ANGLES USED. SMALL ERROR => SMALL IMPACT, EXCEPT FOR TREATENT WITH SMALL MARGINS, E.G. STEREOTACTIC TREATMENTS, WHERE EVEN A SMALL ANGLE DEVIATION COULD HAVE AN IMPACT.
NO EVENT OCCURRED. NO PATIENTS WERE AFFECTED. THERE IS ONLY A POTENTIAL HEALTH HAZARD DISCOVERED BY THE MANUFACTURER. THERE ARE THREE BUGS RELATED TO PITCH AND ROLL PARAMETERS IN REPORTS FOR 6D COUCH: A) BUG 283056: [NON-IEC] DISPLAYED FOR COUCH PITCH AND COUCH ROLL IN HEADER AND BEAM DATA SECTION IN REPORT COUCH PITCH AND ROLL PARAMETERS ARE ALWAYS DEFINED ACCORDING TO IEC61217 STANDARD IN RAYSTATION. HOWEVER, IF THE MACHINE IS DEFINED IN RAYPHYSICS WITH NON-IEC FOR COUCH ANGLE, THE REPORT WILL SHOW UNITS FOR PITCH/ROLL AS "NON-IEC". B) BUG 281806: COUCH ROLL PARAMETER IS NOT SET CORRECTLY FOR SETUP BEAMS IN PLAN REPORTS C) BUG 282677: NO CONVERTER SHOULD BE USED FOR COUCH PITCH/ROLL PARAMETERS IN REPORT SETUP BEAMS TABLE FOR SETUP BEAMS, COUCH ROLL WILL ALWAYS BE INCORRECTLY ASSIGNED THE VALUE OF COUCH ANGLE IN THE PLAN REPORT. IF NON-IEC DEFINITION IS USED FOR COUCH ANGLE, COUCH PITCH WILL ALSO BE INCORRECT.
FOLLOW-UP OF REPORT RSL: MDR 3007774465-2023-00015 68727 RS TREATMENT PLAN REPORT RSL. NO EVENT OCCURRED. NO PATIENTS WERE AFFECTED. THERE IS ONLY A POTENTIAL HEALTH HAZARD DISCOVERED BY THE MANUFACTURER. THERE ARE THREE BUGS RELATED TO PITCH AND ROLL PARAMETERS IN REPORTS FOR 6D COUCH: A) BUG 283056: [NON-IEC] DISPLAYED FOR COUCH PITCH AND COUCH ROLL IN HEADER AND BEAM DATA SECTION IN REPORT COUCH PITCH AND ROLL PARAMETERS ARE ALWAYS DEFINED ACCORDING TO IEC61217 STANDARD IN RAYSTATION. HOWEVER, IF THE MACHINE IS DEFINED IN RAYPHYSICS WITH NON-IEC FOR COUCH ANGLE, THE REPORT WILL SHOW UNITS FOR PITCH/ROLL AS "NON-IEC". B) BUG 281806: COUCH ROLL PARAMETER IS NOT SET CORRECTLY FOR SETUP BEAMS IN PLAN REPORTS C) BUG 282677: NO CONVERTER SHOULD BE USED FOR COUCH PITCH/ROLL PARAMETERS IN REPORT SETUP BEAMS TABLE. FOR SETUP BEAMS, COUCH ROLL WILL ALWAYS BE INCORRECTLY ASSIGNED THE VALUE OF COUCH ANGLE IN THE PLAN REPORT. IF NON-IEC DEFINITION IS USED FOR COUCH ANGLE, COUCH PITCH WILL ALSO BE INCORRECT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 707084 | RAYSTATION | RADIATION THERAPY TREATNEBT PLANNING SYSTEM | MUJ | RAYSEARCH LABORATORIES AB (PUBL) | RAYSTATION 4-10A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |