DAVINCI XI
Report
- Report Number
- 2955842-2023-11081
- Event Type
- Injury
- Date Received
- March 6, 2023
- Date of Event
- February 9, 2023
- Report Date
- February 9, 2023
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874110744
- PMA / PMN Number
- K131861
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
BASED ON THE CURRENT INFORMATION PROVIDED, THE SURGEON CONVERTED THE PROCEDURE TO OPEN SURGERY DUE TO NOT WANTING TO DELAY THE PROCEDURE FOR TROUBLESHOOTING. AFTER THE CASE WAS COMPLETED, THE CUSTOMER WAS ABLE TO RESOLVE THE ISSUE OF NO IMAGE IN THE RIGHT EYE OF THE SURGEON SIDE CONSOLE BY FOLLOWING THE TROUBLESHOOTING STEPS PROVIDED BY INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT. THE SYSTEM IS IN WORKING WITH NO ISSUES. NO SITE VISIT WAS REQUIRED. IF ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP MDR WILL BE SUBMITTED. THIS EVENT IS BEING REPORTED DUE TO THE FOLLOWING CONCLUSION: DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THERE WAS NO IMAGE IN THE RIGHT EYE OF THE SURGEON SIDE CONSOLE (SSC). THE SURGEON DID NOT WANT TO DELAY THE PROCEDURE FOR TROUBLESHOOTING AND ELECTED TO CONVERT THE PROCEDURE TO OPEN SURGERY.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THERE WAS NO RIGHT EYE IMAGE IN THE SURGEON SIDE CONSOLE (SSC). THE INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) VIEWED THE LIVE LOGS AND FOUND A SINGLE 45308 ERROR AND BLUE FIBER RELATED ERRORS. THE TSE RECOMMENDED TURNING THE SYSTEM OFF AND VERIFYING THE BLUE FIBER CONNECTION BETWEEN THE SSC AND VISION SIDE CART (VSC), BUT THE SURGEON DID NOT WANT TO PERFORM A POWER CYCLE. AFTER THE SURGEON HAD CONVERTED THE CASE TO OPEN SURGERY, THE CUSTOMER HARD POWER CYCLED THE SSC AND RESEATED THE BLUE FIBER CABLE AT THE SSC AND VSC. THE SYSTEM POWERED BACK ON WITH NO ISSUES. THE PROCEDURE WAS CONVERTED TO OPEN SURGERY WITH NO REPORTED INJURY. ON (B)(6)2023, ISI FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING INFORMATION: THE PROCEDURE WAS CONVERTED TO OPEN SURGERY WITHOUT ANY INJURY OR HARM TO THE PATIENT. THE SURGEON CONVERTED TO OPEN SURGERY BECAUSE OF THE RIGHT EYE VISION LOSS. THE SURGEON DID NOT WANT TO DELAY THE PROCEDURE FOR TROUBLESHOOTING AND INSTEAD CHOSE TO CONVERT TO OPEN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1631039 | DAVINCI XI | SURGEON SIDE CONSOLE, SMART PEDALS | NAY | INTUITIVE SURGICAL, INC | 380677-23 | N/A | 00886874110744 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES |