FDA Adverse Event Malfunction Summary report: N

TRIAGE TOX DRUG SCREEN WITH MTD

MDR report key: 1648917 · Received March 26, 2010

Report

Report Number
2027969-2010-00365
Event Type
Malfunction
Date Received
March 26, 2010
Date of Event
March 3, 2010
Report Date
March 26, 2010
Manufacturer
BIOSITE INCORPORATED
Product Code
DJR
PMA / PMN Number
K060791
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION PENDING.

Description of Event or Problem · 1

CUSTOMER REPORTED A FALSE NEGATIVE APAP RESULT ON TRIAGE TOX DRUG SCREEN WITH MTD. CUSTOMER RAN FOUR TESTS ON ONE PT URINE SAMPLE AND OBTAINED THE FOLLOWING RESULTS: W45876B: +APAP FRESH URINE. W46864B: -APAP URINE WAS COLD AFTER REFRIGERATION. W45876B: +APAP URINE AT ROOM TEMP > 30 MIN. W46864B: +APAP URINE AT ROOM TEMP > 30 MIN. CUSTOMER USED TWO DIFFERENT STRIP LOTS AND HAD THE SAME RESULTS ON THREE DIFFERENT METERS (42699, 40679 AND 47056). QC SAMPLES TESTED ON BOTH LOTS OF STRIPS WITH CORRECT RESULTS. +APAP IS REPORTED FOR THIS PT. TECH SUPPORT CHECKED WITH THE TECH WHO RAN THE TEST AND SHE RAN THE URINE RIGHT FROM THE REFRIGERATOR (THE NEGATIVE RESULT). SHE WAITED ABOUT A HALF HOUR AND REPEATED, ALL RESULTS VERIFIED.