FDA Adverse Event Malfunction Summary report: N

REVOLUTION DSP 25G, CURVED SCISSORS

MDR report key: 1648722 · Received March 25, 2010

Report

Report Number
3003398873-2010-00003
Event Type
Malfunction
Date Received
March 25, 2010
Date of Event
February 17, 2010
Report Date
February 21, 2010
Manufacturer
ALCON GRIESHABER AG
Product Code
HNF
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE RETURNED PRODUCT WAS VISUALLY INSPECTED. IT WAS FOUND THAT ONE OF THE BLADES OF THE 25GA CURVED SCISSORS IS BROKEN. THE BREAKAGE INDICATES THE PRESENCE OF STRESS CRACK CORROSION. DURING FORMING OF THE INSERTS STRESS CRACKS CAN BE INTRODUCED, IF THE DUCTILITY OF THE MATERIAL IS INSUFFICIENT. SUCH STRESS CRACKS CAN LEAD TO STRESS CRACK CORROSION OVER A LONGER TIME PERIOD. THIS COMPLAINT IS THE FIRST ONE AFTER THE IMPROVEMENT OF THE HARDENING PROCESS. IT IS THE FIRST COMPLAINT AFTER (B)(4) MFG DEVICES. IN FACT, IT IS NOT POSSIBLE TO AVOID SINGLE CASES OF CORROSION DURING THE MFG PROCESSES BECAUSE OF POTENTIAL INHOMOGENEOUS OF THE RAW MATERIAL. THEREFORE, NO CORRECTIVE ACTIONS WILL BE PERFORMED AND THE ISSUE IS REGARDED AS A SINGLE EVENT. (B)(4).

Description of Event or Problem · 1

MALFUNCTION: BROKEN SCISSORS. THE SURGEON REPORTED, WHEN THE SCISSORS WERE INSERTED THROUGH THE TROCAR CANNULA INTO THE EYE, THE CROSS ACTION CUTTING WAS NOT OBSERVED. THE SURGEON NOTED UPON EXTRACTION OF THE SCISSORS, ONE OF THE BLADES WAS BROKEN. THERE WAS NO IMPACT TO THE PT REPORTED. ADDITIONAL INFO HAS BEEN REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 REVOLUTION DSP 25G, CURVED SCISSORS SCISSORS, OPHTHALMIC HNF ALCON GRIESHABER AG 705.52 F73210

Patients

Seq Age Sex Outcome Treatment
1