FDA Adverse Event Injury Summary report: N

HAHN TAPERED IMPLANT Ø3.0 X 11.5 MM

MDR report key: 16485186 · Received March 6, 2023

Report

Report Number
3011649314-2023-00115
Event Type
Injury
Date Received
March 6, 2023
Date of Event
December 13, 2022
Report Date
September 9, 2024
Manufacturer
PRISMATIK DENTALCRAFT, INC.
Product Code
DZE
PMA / PMN Number
K143353
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RI, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CAPA 2023 - 006 THE DEVICE HAS NOT BEEN RETURNED. HOWEVER, THE DEVICE INVESTIGATION HAS BEEN COMPLETED AND THE RESULTS ARE AS FOLLOWS: DHR RESULTS THE DHR WAS REVIEWED FOR HAHN TAPERED IMPLANT LOT# 6113521 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE PART MET ALL THE CRITERIA CALLED FOR IN THE PRODUCTION ROUTER. STOCK PRODUCT REVIEWED RESULTS A REVIEW OF STOCK PRODUCT WAS PERFORMED FOR HAHN TAPERED IMPLANT LOT# 6113521 AND FOUND NO ADDITIONAL PRODUCT IN STOCK. INVESTIGATION METHODS/RESULTS THE DEVICE WAS RETURNED BUT NOT IN ORIGINAL PACKAGE. THE IMPLANT WAS VERIFIED TO BE A HAHN TAPERED IMPLANT Ø3.0 X 11.5 MM (70-1154-IMP0001) USING RADIOGRAPHIC TEMPLATE (PK-209-062515). THERE WAS NO DEFECT OR NON-CONFORMITY OBSERVED AND THE THREADS WERE INTACT. BONE DEBRIS WAS OBSERVED IN THE THREADING OF THE IMPLANT. ROOT CAUSE "FAILURE TO OSSEOINTEGRATE" IS A COMMON COMPLAINT IN REGARDS TO IMPLANT FAILURE. THIS OCCURS WHEN THE PATIENT'S BONE DOES NOT INTEGRATE WITH THE IMPLANT SURFACE. THE POSSIBLE RESPONSES TO THIS COMPLAINT COULD BE ATTRIBUTED TO VARIOUS CAUSES. ALTHOUGH THE ROOT CAUSE FOR FAILURE TO OSSEOINTEGRATE IS INCONCLUSIVE AND SPECIFIC TO EACH CASE, PROBABLE CAUSES COULD BE THE LOSS OF PRIMARY STABILITY AT THE OSTEOTOMY SITE DUE TO INSUFFICIENT BONE OR POOR BONE QUALITY; EITHER THE BONE WAS TOO SOFT OR THE OPERATOR ERRED IN CREATING AN OSTEOTOMY BIGGER THAN THE SIZE OF THE IMPLANT DIAMETER. PREMATURE LOADING, PATIENT'S HEALTH, PERI-IMPLANTITIS, SMOKING, AND LACK OF ORAL HYGIENE MAY ALSO BE CONTRIBUTING FACTORS. PER THE REPORTED INFORMATION, IFU 570 REV 3.0 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE CONTRAINDICATIONS SECTION: "HAHN TAPERED IMPLANTS SHOULD NOT BE PLACED IN PATIENTS DISCOVERED TO BE MEDICALLY UNFIT FOR THE INTENDED TREATMENT. PRIOR TO CLINICAL INTERVENTION, PROSPECTIVE PATIENTS MUST BE THOROUGHLY EVALUATED FOR ALL KNOWN RISK FACTORS AND CONDITIONS RELATED TO ORAL SURGICAL PROCEDURES AND SUBSEQUENT HEALING. CONTRAINDICATIONS INCLUDE BUT ARE NOT LIMITED TO: VASCULAR CONDITIONS, UNCONTROLLED DIABETES, CLOTTING DISORDERS, ANTICOAGULANT THERAPY, METABOLIC BONE DISEASE, CHEMOTHERAPY OR RADIATION THERAPY, CHRONIC PERIODONTAL INFLAMMATION, INSUFFICIENT SOFT TISSUE COVERAGE, METABOLIC OR SYSTEMIC DISORDERS ASSOCIATED WITH WOUND AND/OR BONE HEALING, USE OF PHARMACEUTICALS THAT INHIBIT OR ALTER NATURAL BONE REMODELING, ANY DISORDERS WHICH INHIBIT A PATIENT'S ABILITY TO MAINTAIN ADEQUATE DAILY ORAL HYGIENE, UNCONTROLLED PARAFUNCTIONAL HABITS, INSUFFICIENT HEIGHT AND/OR WIDTH OF BONE, AND INSUFFICIENT INTERARCH SPACE." IFU 570 REV 3.0 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN PRECAUTION SECTION SURGICAL PROCEDURES: "MINIMIZING TISSUE DAMAGE IS CRUCIAL TO SUCCESSFUL IMPLANT OSSEOINTEGRATION. CARE SHOULD BE TAKEN TO ELIMINATE SOURCES OF INFECTION, CONTAMINANTS, SURGICAL AND THERMAL TRAUMA. THE RISK OF OSSEOINTEGRATION FAILURE INCREASES AS TISSUE TRAUMA INCREASES. ALL DRILLING PROCEDURES SHOULD BE PERFORMED AT 2000 RPM OR LESS UNDER CONTINUAL AND COPIOUS IRRIGATION. ALL SURGICAL INSTRUMENTS USED MUST BE IN GOOD CONDITION AND SHOULD BE USED CAREFULLY TO AVOID DAMAGE TO IMPLANTS OR OTHER COMPONENTS. IMPLANTS SHOULD BE PLACED WITH SUFFICIENT STABILITY; HOWEVER, EXCESSIVE INSERTION TORQUE MAY RESULT IN IMPLANT FRACTURE, OR FRACTURE OR NECROSIS OF THE IMPLANT SITE. THE PROPER SURGICAL PROTOCOL SHOULD BE STRICTLY ADHERED TO." IFU 570 REV 3.0 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: "ABSOLUTE SUCCESS CANNOT BE GUARANTEED. FACTORS SUCH AS INFECTION, DISEASE AND INADEQUATE BONE QUALITY AND/OR QUANTITY CAN RESULT IN OSSEOINTEGRATION FAILURES FOLLOWING SURGERY OR INITIAL OSSEOINTEGRATION." IFU 570 REV 3.0 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: "THE IMPLANT SITE SHOULD BE INSPECTED FOR ADEQUATE BONE BY RADIOGRAPHS, PALPATIONS AND VISUAL EXAMINATION. DETERMINE THE LOCATION OF NERVES AND OTHER VITAL STRUCTURES AND THEIR PROXIMITY TO THE IMPLANT SITE BEFORE ANY DRILLING TO AVOID POTENTIAL INJURY, SUCH AS PERMANENT NUMBNESS TO THE LOWER LIP AND CHIN." THIS COMPLAINT WILL BE KEPT ON RECORD FOR TRACK AND TRENDING PURPOSES.

Additional Manufacturer Narrative · 0

THE DEVICE HAS NOT BEEN RETURNED. IF/ WHEN THE DEVICE IS RETURNED AN INVESTIGATION WILL BE CARRIED OUT AND A SUPPLEMENTAL REPORT WILL BE SUBMITTED. WEIGHT: THE PATIENT'S WEIGHT IS NOT RECORDED AT THE TIME OF THE VISIT.

Additional Manufacturer Narrative · 0

CAPA 23-006. MANUFACTURER REFERENCE: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT THE HAHN TAPERED IMPLANT FAILED TO INTEGRATE. THE PATIENT'S BONE TYPE IS II AND THEIR ORAL HYGIENE IS LISTED AS GOOD. THERE IS NO MEDICAL OR DENTAL HISTORY PRIOR TO IMPLANT PLACEMENT. THE PATIENT PRESENTED ON (B)(6)2022 FOR A PRIMARY PROCEDURE ON TOOTH #13. THE PATIENT RETURNED ON (B)(6)2022, AT THE SECOND STAGE SURGERY WITH COMPLAINTS OF PAIN. UPON EXAMINATION, THE PROVIDER NOTED ABNORMAL BONE LOSS AROUND THE IMPLANT. IT WAS DETERMINED THAT THE IMPLANT FAILED TO INTEGRATE AND DEVICE WAS REMOVED. BONE GRAFTING WAS REQUIRED. THE IMPLANT WILL BE REPLACED WITH THE SAME PRODUCT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
708323 HAHN TAPERED IMPLANT Ø3.0 X 11.5 MM HAHN TAPERED IMPLANT DZE PRISMATIK DENTALCRAFT, INC. 70-1154-IMP0001 6113521

Patients

Seq Age Sex Outcome Treatment
1 40 YR Female Required Intervention