FDA Adverse Event Injury Summary report: N

PITT EASY ABUTMENT

MDR report key: 1648440 · Received April 1, 2010

Report

Report Number
9611820-2010-00004
Event Type
Injury
Date Received
April 1, 2010
Report Date
March 2, 2010
Manufacturer
SYBRON IMPLANT SOLUTIONS GMBH
Product Code
NHA
PMA / PMN Number
K053242
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

A DOCTOR REPORTED THAT ONE PITT-EASY ABUTMENT SCREW FRACTURED WHILE HE WAS RATCHETING IT INTO PLACE. THE DOCTOR STATED THAT THE RACHET HE USED WAS NOT WORKING PROPERLY AND HAD NOT BEEN PROPERLY CALIBRATED. HE ALSO STATED THAT THE RACHET WAS NOT MANUFACTURED BY SYBRON IMPLANT SOLUTIONS (B) (4). THE ABUTMENT WILL BE REPLACED USING THE PROPER RACHET. NO PATIENT INFORMATION WAS PROVIDED AND THE PRODUCT HAS NOT YET BEEN RETURNED TO SYBRON IMPLANT SOLUTIONS (B) (4) FOR EVALUATION. WHEN AN EVALUATION OF THE FRACTURED ABUTMENT HAS BEEN CONDUCTED A FOLLOW-UP REPORT WILL BE SUBMITTED. DEVICE WAS NOT RETURNED TO MANUFACTURER.

Description of Event or Problem · 1

ON (B) (6) 2010 A DOCTOR REPORTED THAT ONE PITT-EASY ABUTMENT SCREW FRACTURED WHILE HE WAS RATCHETING IT INTO PLACE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PITT EASY ABUTMENT ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS NHA SYBRON IMPLANT SOLUTIONS GMBH 949201

Patients

Seq Age Sex Outcome Treatment
1 Other| R