PENTAX
Report
- Report Number
- 9610877-2023-52018
- Event Type
- Malfunction
- Date Received
- March 6, 2023
- Date of Event
- February 16, 2023
- Report Date
- March 6, 2023
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- EOQ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EB15-J10 IS AVAILABLE IN THE USA WITH A 510K NUMBER K200678. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE DISTAL END ASSY W/CCD MODULE FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE DISTAL END ASSY W/CCD MODULE. IN ADDITION, WE CONFIRMED THAT THE BENDING RUBBER PERFORATED, THE INSERTION FLEXIBLE TUBE (IFT) CRUSHED, THE LG PRONG DEFORMED, AND THE REMOTE CONTROL BUTTON(1) MALFUNCTION; HOWEVER, THEY ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586 (IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (FLUID DAMAGE).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1513206 | PENTAX | VIDEO BRONCHOSCOPE | EOQ | HOYA CORPORATION PENTAX TOKYO OFFICE | EB11-J10 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |