FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 16483796 · Received March 6, 2023

Report

Report Number
9610877-2023-52018
Event Type
Malfunction
Date Received
March 6, 2023
Date of Event
February 16, 2023
Report Date
March 6, 2023
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
EOQ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EB15-J10 IS AVAILABLE IN THE USA WITH A 510K NUMBER K200678. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE DISTAL END ASSY W/CCD MODULE FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE DISTAL END ASSY W/CCD MODULE. IN ADDITION, WE CONFIRMED THAT THE BENDING RUBBER PERFORATED, THE INSERTION FLEXIBLE TUBE (IFT) CRUSHED, THE LG PRONG DEFORMED, AND THE REMOTE CONTROL BUTTON(1) MALFUNCTION; HOWEVER, THEY ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586 (IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (FLUID DAMAGE).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1513206 PENTAX VIDEO BRONCHOSCOPE EOQ HOYA CORPORATION PENTAX TOKYO OFFICE EB11-J10

Patients

Seq Age Sex Outcome Treatment
1 Unknown