IMPL TAPERED SP 4.8MM 8MM OCTAGON
Report
- Report Number
- 0002023141-2023-00714
- Event Type
- Injury
- Date Received
- March 3, 2023
- Date of Event
- January 11, 2023
- Report Date
- July 25, 2023
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024017597
- PMA / PMN Number
- K082639
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PL
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ZIMMER BIOMET COMPLAINT NUMBER (B)(4). PATIENT AGE AND DATE OF BIRTH UNKNOWN / NOT PROVIDED. PATIENT WEIGHT UNKNOWN / NOT PROVIDED . LOT/SERIAL # UNKNOWN / NOT PROVIDED . DEVICE EXPIRATION DATE UNKNOWN / NOT PROVIDED. DEVICE UDI NUMBER UNKNOWN / NOT PROVIDED. K011245, K002188. DEVICE MANUFACTURER DATE UNKNOWN / NOT PROVIDED. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.
ZIMMER BIOMET COMPLAINT NUMBER (B)(4). G4: PREMARKET IDENTIFICATION K082639/K011245/K002188. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 2022051095. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. STERILIZATION RECORD (B)(4) WAS REVIEWED AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 2022051095 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: ¿DENTAL: MEDICAL: PERI-IMPLANTITIS¿. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.
THE FOLLOWING SECTIONS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT D4: LOT/SERIAL # D4: DEVICE EXPIRATION DATE D4: DEVICE UDI NUMBER D9: DEVICE AVAILABILITY G3: DATE RECEIVED BY MANUFACTURER G6: CHECKED "FOLLOW-UP" H2: CHECKED FOLLOW-UP TYPE H4: DEVICE MANUFACTURER DATE H6: ENTERED EVALUATION CODES H10: ADDED MANUFACTURER NARRATIVE.
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE 20 WAS REMOVED DUE TO PERI-IMPLANTITIS. EDEMA AND PAIN WERE REPORTED AS A RESULT OF THE EVENT. NO DELAY, PROCEDURE WAS NOT COMPLETED AND PATIENT WILL NOT HAVE TO RETURN.
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE 35 WAS REMOVED DUE TO PERI-IMPLANTITIS. EDEMA AND PAIN WERE REPORTED AS A RESULT OF THE EVENT. NO DELAY, PROCEDURE WAS NOT COMPLETED AND PATIENT WILL NOT HAVE TO RETURN.
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2001423 | IMPL TAPERED SP 4.8MM 8MM OCTAGON | DENTAL IMPLANT | DZE | ZIMMER DENTAL | SPWB8 | 2022051095 | 00889024017597 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Prefer Not To Disclose | Required Intervention |