FDA Adverse Event Malfunction Summary report: N

ICELOCK 621 RATCHET

MDR report key: 16478029 · Received March 3, 2023

Report

Report Number
3003764610-2023-00007
Event Type
Malfunction
Date Received
March 3, 2023
Date of Event
January 30, 2023
Report Date
March 15, 2023
Manufacturer
OSSUR HF
Product Code
ISH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PREMATURE PRODUCT WEAR RESULTED IN LOCK NOT PROVIDING SECURE SUSPENSION. PRODUCT HAD BEEN IN USE FOR 4 MONTHS. USER DEVELOPED SKIN REACTION ON THE RESIDUAL LIMB FROM HOBBLING AROUND WHILE ON HOLIDAY. PLAY AND PREMATURE LOCKING PLATE WEAR HAS BEEN IDENTIFIED AS LIKELY ROOT CAUSE OF FAILURE. CAPA IS IN PROGRESS TO ADDRESS ISSUES WITH PREMATURE WEAR OF THE LOCK. PREMATURE WEAR CAN LEAD TO SERIOUS INJURY, BUT ISSUE IS EXPECTED TO BE DISCOVERED BY THE USER BEFORE BECOMING HAZARDOUS AS THE WEAR OCCURS GRADUALLY OVER TIME AND CHANGE IN USE OF THE LOCKING MECHANISM IS LIKELY TO BE NOTICED BY THE USER. INSTRUCTIONS FOR USE INDICATES A WARNING FOR CHANGE OR LOSS IN DEVICE FUNCTIONALITY, AND TO CONTACT A CLINICIAN WHEN THIS OCCURS. THE MAJORITY OF CLAIMS REGARDING THIS ISSUE WERE NOT ASSOCIATED WITH AN INCIDENT, BUT DISCOVERED IN A TIMELY MANNER.

Description of Event or Problem · 0

LOCK WAS CLICKING FOR A COUPLE OF DAYS, THEN ONE DAY STOPPED ENGAGING WITH THE PIN, PIN RELEASED FROM SOCKET AND WOULD NOT STAY ON. CUSTOMER REPORTED THAT PATIENT HAD TO HOBBLE AROUND AND GOT A BLISTER AT THE BOTTOM OF THE RESIDUAL LIMB.

Description of Event or Problem · 0

LOCK WAS CLICKING FOR A COUPLE OF DAYS, THEN ONE DAY STOPPED ENGAGING WITH THE PIN, PIN RELEASED FROM SOCKET AND WOULD NOT STAY ON. CUSTOMER REPORTED THAT PATIENT HAD TO HOBBLE AROUND AND GOT A BLISTER AT THE BOTTOM OF THE RESIDUAL LIMB.

Description of Event or Problem · 0

LOCK WAS CLICKING FOR A COUPLE OF DAYS, THEN ONE DAY STOPPED ENGAGING WITH THE PIN, PIN RELEASED FROM SOCKET AND WOULD NOT STAY ON. CUSTOMER REPORTED THAT PATIENT HAD TO HOBBLE AROUND AND GOT A BLISTER AT THE BOTTOM OF THE RESIDUAL LIMB.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1732023 ICELOCK 621 RATCHET PROSTHETIC LOCK ISH OSSUR HF L-621000

Patients

Seq Age Sex Outcome Treatment
1 Unknown Congenital Anomaly