FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 10MM

MDR report key: 16473408 · Received March 2, 2023

Report

Report Number
0002023141-2023-00705
Event Type
Injury
Date Received
March 2, 2023
Date of Event
January 3, 2023
Report Date
June 12, 2023
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024020047
PMA / PMN Number
K061410
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). PATIENT WEIGHT UNKNOWN / NOT PROVIDED. PMA/ 510K: /K011028/K013227.

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). THE FOLLOWING FIELDS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT . B5: DESCRIBE EVENT OR PROBLEM. G3: DATE RECEIVED BY MANUFACTURER. G6: TYPE OF REPORT. H1: TYPE OF REPORTABLE EVENT. H2: FOLLOW UP TYPE. H3: DEVICE EVALUATED BY MANUFACTURER. H6: ADVERSE EVENT PROBLEM. H10: ADDITIONAL NARRATIVE. ONE IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 10MM (TSVWB10) WAS RETURNED FOR INVESTIGATION. VISUAL EVALUATION OF THE AS RETURNED IMPLANT IDENTIFIED SIGNS OF USE. THERE WAS DAMAGE FROM THE REMOVAL PROCESS. DAMAGE SEEN ON THE EXTERNAL THREADS AND COLLAR. THE IMPLANTS DRIVE FEATURE WAS DAMAGED. ZIMVIE IS NOT RESPONSIBLE FOR ANY DAMAGE CAUSED BY THE CLINICIAN'S REMOVAL OF THE DEVICE. FUNCTIONAL TESTING COULD NOT BE PERFORMED SINCE THE PRODUCT WAS DAMAGED. PRE-EXISTING CONDITION NOTED ON THE PER WAS UNKNOWN BONE DENSITY. THE REPORTED DEVICE HAD BEEN PLACED ON TOOTH #46 (FDI) FOR APPROXIMATELY 2 YEARS. DHR REVIEW FOR THE LOT NUMBER (1240410) HAD REVEALED NO DEVIATIONS NOR NON-CONFORMANCES WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. PRODUCTS WAS CONFORMING AT THE TIME IT LEFT ZIMVIE. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1240410) FOR SIMILAR EVENTS (COMPLAINT CATEGORY KEYWORDS: FUNCTIONAL: DAMAGED DRIVE FEATURE) AND NO OTHER COMPLAINT WAS IDENTIFIED. THEREFORE, BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID OCCUR, AND THE REPORTED EVENT WAS CONFIRMED.

Description of Event or Problem · 0

DOCTOR REPORTED FLARED COLLAR AND WAS REMOVED. NO OTHER IMPLANT HAS BEEN PLACED FOR THE MOMENT. #46.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2002939 IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 10MM DENTAL IMPLANT DZE ZIMMER DENTAL TSVWB10 1240410 00889024020047

Patients

Seq Age Sex Outcome Treatment
1 67 YR Female Required Intervention