FDA Adverse Event Malfunction Summary report: N

TREO ABDOMINAL STENT-GRAFT SYSTEM

MDR report key: 16470008 · Received March 2, 2023

Report

Report Number
2247858-2023-00063
Event Type
Malfunction
Date Received
March 2, 2023
Date of Event
February 2, 2023
Report Date
July 14, 2023
Manufacturer
BOLTON MEDICAL, INC.
Product Code
MIH
PMA / PMN Number
P190015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

BOLTON MEDICAL, INC. (D/B/A TERUMO AORTIC), HEREIN KNOWN AS THE "COMPANY", IS SUBMITTING THIS REPORT PURSUANT TO 21 CFR PART 803, HAS MADE REASONABLE EFFORTS TO OBTAIN COMPLETE INFORMATION, AND HAS PROVIDED AS MUCH AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT IS BASED ON INFORMATION OBTAINED BY THE COMPANY, WHICH MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE REGULATIONS. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN, UNAVAILABLE, OR NOT APPLICABLE. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY ANYONE THAT THE DEVICE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE A LEGAL ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY DEFECTS OR HAS MALFUNCTIONED, AS DEFINED FROM A LEGAL STANDPOINT. THESE WORDS ARE INCLUDED IN THE REPORT AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE CERTAIN REGULATIONS, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

PATIENT HAD EXISTING TRI-VASCULAR OVATION GRAFT IN PLACE WITH TYPE 1A LEAK. SURGEON DECIDED TO CUFF UP TO RIGHT RENAL ARTERY. ONCE HE DEPLOYED GRAFT, AND HAD SEPARATED LEAVE BEHIND SHEATH FROM DEPLOYMENT DEVICE, THE NOSE CONE CLEARLY SEPARATED FROM SYSTEM. SURGEON REPORTED FEELING NO RESISTANCE, AND WE ONLY SAW IT HAPPENED BY WATCHING FLUOROSCOPY. NOSE CONE REMAINED ON DOUBLE CURVE LUNDERQUIST WIRE; SURGEON WAS ABLE TO SUCCESSFULLY SNARE THE NOSE CONE AND BRING IT BACK INTO 20F DRYSEAL SHEATH. PATIENT OUTCOME - "PATIENT IS OKAY."

Description of Event or Problem · 0

PATIENT HAD EXISTING TRI-VASCULAR OVATION GRAFT IN PLACE WITH TYPE 1A LEAK. SURGEON DECIDED TO CUFF UP TO RIGHT RENAL ARTERY. ONCE HE DEPLOYED GRAFT, AND HAD SEPARATED LEAVE BEHIND SHEATH FROM DEPLOYMENT DEVICE, THE NOSE CONE CLEARLY SEPARATED FROM SYSTEM. SURGEON REPORTED FEELING NO RESISTANCE, AND WE ONLY SAW IT HAPPENED BY WATCHING FLUOROSCOPY. NOSE CONE REMAINED ON DOUBLE CURVE LUNDERQUIST WIRE; SURGEON WAS ABLE TO SUCCESSFULLY SNARE THE NOSE CONE AND BRING IT BACK INTO 20F DRYSEAL SHEATH. PATIENT OUTCOME - "PATIENT IS OKAY."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
516681 TREO ABDOMINAL STENT-GRAFT SYSTEM STENT, ENDOVASCULAR GRAFT, AORTIC MIH BOLTON MEDICAL, INC. 2209160040

Patients

Seq Age Sex Outcome Treatment
1 79 YR Female Required Intervention