FDA Adverse Event Malfunction Summary report: N

5.5MM HEALIX ADVANCE¿ BR ANCHOR WITH DYNACORD¿ SUTURE (BLE, WHI/BLU/GRN, WH/BLK)

MDR report key: 16461838 · Received March 1, 2023

Report

Report Number
1221934-2023-00927
Event Type
Malfunction
Date Received
March 1, 2023
Date of Event
February 7, 2023
Manufacturer
MEDOS INTERNATIONAL SARL
Product Code
MAI
UDI-DI
10886705029457
PMA / PMN Number
K173859
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED IN PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY MITEK OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY MITEK, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. (B)(4). THE LOT NUMBER WAS UNKNOWN. THE LOT NUMBER WAS UNKNOWN; THEREFORE, THE EXPIRATION DATE, MANUFACTURING SITE NAME AND DEVICE MANUFACTURE DATE WERE UNKNOWN. INITIAL REPORTER OCCUPATION: REPORTER IS A J&J SALES REPRESENTATIVE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED IN PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY MITEK OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY MITEK, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILALE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. H10 ADDITIONAL NARRATIVE: INVESTIGATION SUMMARY - THE COMPLAINT DEVICE IS NOT BEING RETURNED, IT WAS DISCARDED BY THE CUSTOMER, THEREFORE UNAVAILABLE FOR A PHYSICAL EVALUATION. SINCE THE COMPLAINT DEVICE WAS NOT RETURNED, WE CANNOT DETERMINE A ROOT CAUSE FOR THE REPORTED FAILURE. IF ADDITIONAL INFORMATION IS RECEIVED IN THE FUTURE, WE WILL REOPEN THE COMPLAINT AND PERFORM THE INVESTIGATION AS APPROPRIATE. NO LOT NUMBERS WERE SUPPLIED WHICH PRECLUDES CONDUCTING A DHR REVIEW OR A LOT OF SPECIFIC SEARCH IN THE COMPLAINTS HANDLING SYSTEM. AT THIS POINT IN TIME, NO CORRECTIVE ACTION IS REQUIRED, AND NO FURTHER ACTION IS WARRANTED. HOWEVER, DEPUY SYNTHES MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD.

Description of Event or Problem · 0

IT WAS REPORTED BY THE SALES REP THAT DURING A ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2023, A 5.5MM HEALIX ADVANCE¿ BR ANCHOR WITH DYNACORD¿ SUTURE (BLUE, WHITE/BLUE/GREEN, WHITE/BLACK) DEVICE WAS USED. ACCORDING TO THE REPORT, THE ANCHOR BRIDGE THAT HOLDS ANCHOR IN PLACE BROKE AND RELEASED SUTURES FROM SUTURE ANCHOR WHILE SURGEON WAS TYING KNOTS. IT WAS FURTHER REPORTED THAT THE SURGEON REMOVED THE ANCHOR WITH THE INSERTER AND PLACED A LARGER ANCHOR IN ITS PLACE TO COMPLETE THE PROCEDURE WITH A TWO MINUTE DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1181758 5.5MM HEALIX ADVANCE¿ BR ANCHOR WITH DYNACORD¿ SUTURE (BLE, WHI/BLU/GRN, WH/BLK) SOFT-TISSUE ANCHOR, BIOABSORBABLE MAI MEDOS INTERNATIONAL SARL 222005 10886705029457

Patients

Seq Age Sex Outcome Treatment
1 Unknown