FDA Adverse Event Malfunction Summary report: N

COBAS LITHIUM

MDR report key: 16456723 · Received February 28, 2023

Report

Report Number
1823260-2023-00621
Event Type
Malfunction
Date Received
February 28, 2023
Date of Event
January 30, 2023
Report Date
April 3, 2023
Manufacturer
ROCHE DIAGNOSTICS
Product Code
NDW
UDI-DI
07613336121382
PMA / PMN Number
K063684
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE LAST CALIBRATION PERFORMED ON 29-JAN-2023 WAS WITHIN SPECIFICATIONS. THE QC RECOVERY WAS WITHIN -2 STANDARD DEVIATIONS (SD). THERE IS NO INDICATION OF A PERFORMANCE ISSUE WITH THE REAGENT. THE FIELD APPLICATIONS SPECIALIST (FAS) INSPECTED THE ANALYZER AND FOUND THAT THE ISSUE WAS DUE TO INTERFERENCE WITH THE HDL ASSAY. THE CUSTOMER WORKED WITH FAS TO CONFIGURE AN EXTRA WASH CYCLE. THEY PERFORMED OPERATIONAL TESTS AND QC WITH ACCEPTABLE RESULTS. AFTER THIS INTERVENTION, NO FURTHER ISSUES WERE REPORTED BY THE CUSTOMER.  PRODUCT LABELING STATES "TO AVOID CARRYOVER BETWEEN TESTS, THE SYSTEM CAN PERFORM SPECIAL WASHES . DOWNLOAD ALL REQUIRED SPECIAL WASHES ( MANUFACTURER RULES). YOU CAN PROGRAM ADDITIONAL SPECIAL WASHES MANUALLY ( USER RULES)." THE INVESTIGATION DETERMINED THE SERVICE ACTIONS RESOLVED THE ISSUE.

Description of Event or Problem · 0

THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR TWO PATIENT SAMPLES TESTED WITH LI LITHIUM ON A COBAS PRO C 503 ANALYTICAL UNIT SERIAL NUMBER (B)(4). THE INITIAL VALUES WERE REPORTED OUTSIDE OF THE LABORATORY. DUE TO PREVIOUS ISSUES, THE SAMPLES WERE REPEATED ON A SECOND ANALYZER ON 03-FEB-2023 AND THESE VALUES WERE DEEMED CORRECT. THE FIRST SAMPLE INITIALLY RESULTED IN A LITHIUM VALUE OF 1.2 MMOL/L AND REPEATED AS 0.7 MMOL/L. THE SECOND SAMPLE INITIALLY RESULTED IN A LITHIUM VALUE OF 1.9 MMOL/L AND REPEATED AS 0.6 MMOL/L.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1142815 COBAS LITHIUM LITHIUM ASSAY NDW ROCHE DIAGNOSTICS NA 59094201 07613336121382

Patients

Seq Age Sex Outcome Treatment
1 50 YR Female