FDA Adverse Event Malfunction Summary report: N

VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 0°, AUTOCLAVABLE

MDR report key: 16453498 · Received February 28, 2023

Report

Report Number
9610773-2023-00598
Event Type
Malfunction
Date Received
February 28, 2023
Report Date
March 30, 2023
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
HET
UDI-DI
04042761074964
PMA / PMN Number
K111788
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS MEDWATCH IS BEING SUPPLEMENTED WITH ADDITIONAL INFORMATION OBTAINED AND THE MANUFACTURER'S FINAL INVESTIGATION RESULTS. DURING INSPECTION, OSTE IDENTIFIES A GREEN IMAGE AND TRACES IT BACK TO THE R-UNIT. POSSIBLE CAUSES: UNACCEPTABLE TENSION IN THE AREA OF THE "CCD (CHARGED COUPLED DEVICE) W. HOLDER" ASSEMBLY DUE TO USE OF AN ADHESIVE WHICH IS NOT ADAPTED TO THE LOAD / TEMPERATURE COLLECTIVE AND TO AN INSUFFICIENTLY FORMED ADHESIVE LAYER "C-HOLDER TO BUMPER SLEEVE" UNACCEPTABLE TENSION IN THE AREA OF THE "CCD W. HOLDER" ASSEMBLY DUE TO ASYMMETRICAL ALIGNMENT OF THE SPRING TO C-HOLDER BEFORE THE BUMPER SLEEVE IS JOINED PRE-EXISTING DAMAGE TO THE "CCD W. HOLDER" ASSEMBLY DUE TO USING A SUBOPTIMAL ADHESIVE DEVICE¿ A MANUFACTURING AND QUALITY CONTROL REVIEW WAS PERFORMED FOR THE AFFECTED SERIAL NUMBER WITHOUT SHOWING ANY NON-CONFORMITIES OR DEVIATIONS REGARDING THE DESCRIBED ISSUES.

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO OLYMPUS FOR EVALUATION AND REPAIR. THE INVESTIGATION IS STILL IN PROGRESS; THEREFORE, THE ROOT CAUSE OF THE REPORTED DEFECT CANNOT BE DETERMINED AT THIS TIME. HOWEVER, IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED ACCORDINGLY. IN GENERAL, THE CUSTOMER IS REQUIRED TO INSPECT THE DEVICE FOR DEFECTS, CHECK THE FUNCTION OF ALL DEVICES AND HAVE ALTERNATE EQUIPMENT PRIOR TO USE. OLYMPUS WILL CONTINUE TO MONITOR THE FIELD PERFORMANCE OF THIS DEVICE. ADDITIONAL PMA/510(K): K190744.

Description of Event or Problem · 0

IT WAS REPORTED TO OLYMPUS THAT A VIDEO TELESCOPE HAD A GREEN IMAGE. THE REPORTABLE EVENT WAS FOUND DURING INSPECTION OF THE DEVICE. THERE WAS NO INJURY OR HEALTH DAMAGE DUE TO THE REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
388099 VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 0°, AUTOCLAVABLE LAPAROSCOPE HET OLYMPUS WINTER & IBE GMBH WA50040A 04042761074964

Patients

Seq Age Sex Outcome Treatment
1 Unknown