FDA Adverse Event Malfunction Summary report: N

RAYSTATION

MDR report key: 16452872 · Received February 28, 2023

Report

Report Number
3007774465-2023-00002
Event Type
Malfunction
Date Received
February 28, 2023
Date of Event
May 13, 2020
Report Date
March 5, 2025
Manufacturer
RAYSEARCH LABORATORIES AB (PUBL)
Product Code
MUJ
PMA / PMN Number
K200569
Removal / Correction Number
RES 85723
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THERE HAS BEEN NO ADVERSE EVENT REPORTED FOR THE PB DOSE ERROR. HOWEVER, DURING INVESTIGATION OF A REPORTED DISCREPANCY WHEN USING THE MC DOSE ENGINE, A SOFTWARE IMPLEMENTATION ERROR WAS IDENTIFIED. THE CAUSE OF THE SOFTWARE ERROR IS CURRENTLY NOT KNOWN. ROOT CAUSE ANALYSIS IS IN PROGRESS. IN THE EVENT THAT A CLINICAL DECISION IS BASED ON THE INCORRECTLY CALCULATED DOSE, THIS COULD LEAD TO THE APPROVAL OF AN INAPPROPRIATE DOSE PLAN. THIS COULD LEAD TO LOCAL OVER-DOSE FROM ALLOWING TOO HIGH OF A DOSE IN VOLUMES AFFECTED BY SPOTS WITH CENTRAL AXIS OUTSIDE THE EXTERNAL ROI.

Description of Event or Problem · 0

DESCRIPTION OF REPORTED ADVERSE EVENT: NO SERIOUS ADVERSE EVENT REPORTED. AN ISSUE WITH THE PROTON PENCIL BEAM DOSE ENGINE WAS FOUND DURING INVESTIGATION OF A REPORT OF AN UNEXPECTED BEHAVIOUR WHEN USING THE PROTON MONTE CARLO DOSE ENGINE. A SOFTWARE ERROR IN THE PROTON PENCIL BEAM (PB) DOSE ENGINE HAS BEEN FOUND. SPOTS FOR WHICH THE CENTRAL AXIS DOES NOT GO THROUGH THE EXTERNAL ROI (I.E. THE REGION OF INTEREST THAT DEFINES THE VOLUME WHERE DOSE CALCULATION SHALL BE PERFORMED) ARE NOT INCLUDED IN THE PB PROTON DOSE CALCULATION EVEN IF PARTS OF THE SPOT FALL INSIDE THE EXTERNAL ROI. THIS LEADS TO AN UNDERESTIMATION OF THE DOSE THAT WOULD BE DELIVERED IN VOLUMES AFFECTED BY SUCH SPOTS. THE MONTE CARLO (MC) PROTON OPTIMIZATION DOSE ENGINE CONTAINS THE SAME BUG, BUT THE FINAL MC DOSE CALCULATION IS CORRECT. A CUSTOMER REPORTED SIGNIFICANT DISCREPANCY BETWEEN OPTIMIZED AND FINAL MC DOSE. THIS IS NOT A SAFETY ISSUE, BUT INVESTIGATION OF THIS REPORT IS WHAT LEAD TO THE DISCOVERY OF THE SAFETY RELATED PB DOSE ERROR. THERE WERE NO DEATHS OR SERIOUS INJURIES RELATED TO THIS MALFUNCTION.

Description of Event or Problem · 0

FOLLOW-UP OF REPORT RSL: MDR 3007774465-2023-00002 (B)(6) PROTON DOSE FOR SPOTS OUTSIDE EXTERNAL. DESCRIPTION OF REPORTED ADVERSE EVENT: NO SERIOUS ADVERSE EVENT REPORTED. AN ISSUE WITH THE PROTON PENCIL BEAM DOSE ENGINE WAS FOUND DURING INVESTIGATION OF A REPORT OF AN UNEXPECTED BEHAVIOUR WHEN USING THE PROTON MONTE CARLO DOSE ENGINE. A SOFTWARE ERROR IN THE PROTON PENCIL BEAM (PB) DOSE ENGINE HAS BEEN FOUND. SPOTS FOR WHICH THE CENTRAL AXIS DOES NOT GO THROUGH THE EXTERNAL ROI (I.E. THE REGION OF INTEREST THAT DEFINES THE VOLUME WHERE DOSE CALCULATION SHALL BE PERFORMED) ARE NOT INCLUDED IN THE PB PROTON DOSE CALCULATION EVEN IF PARTS OF THE SPOT FALL INSIDE THE EXTERNAL ROI.THIS LEADS TO AN UNDERESTIMATION OF THE DOSE THAT WOULD BE DELIVERED IN VOLUMES AFFECTED BY SUCH SPOTS. THE MONTE CARLO (MC) PROTON OPTIMIZATION DOSE ENGINE CONTAINS THE SAME BUG, BUT THE FINAL MC DOSE CALCULATION IS CORRECT. A CUSTOMER REPORTED SIGNIFICANT DISCREPANCY BETWEEN OPTIMIZED AND FINAL MC DOSE. THIS IS NOT A SAFETY ISSUE, BUT INVESTIGATION OF THIS REPORT IS WHAT LEAD TO THE DISCOVERY OF THE SAFETY RELATED PB DOSE ERROR. THERE WERE NO DEATHS OR SERIOUS INJURIES RELATED TO THIS MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
560878 RAYSTATION RADIATION THERAPY TREATMENT PLANNING SYSTEM MUJ RAYSEARCH LABORATORIES AB (PUBL) RAYSTATION 4, 4.5, 4.7, 5, 6, 7, 8A, 9B, 9A, 9B, 9B SP1

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other