ESPRIT VENTILATOR
Report
- Report Number
- 2031642-2010-00053
- Event Type
- Malfunction
- Date Received
- March 30, 2010
- Date of Event
- March 4, 2010
- Report Date
- March 4, 2010
- Manufacturer
- RESPIRONICS CALIFORNIA, INC.
- Product Code
- CBK
- PMA / PMN Number
- K981072
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
NO LONGER UNDER WARRANTY FOR SERVICE BY RESPIRONICS; BIOMED PERFORMS SERVICE. RESPIRONICS CALIFORNIA INC. SERVICE TECHNICIAN PROVIDED SERVICE ASSISTANCE TO BIOMED.
THE BIOMEDICAL ENGINEER REPORTED THE UNIT ALARMED AND WENT VENT INOP AFTER SOME POWER ISSUES AT THE HOSPITAL. THERE WAS NO PATIENT HARM REPORTED. REVIEW OF THE VENTILATOR DIAGNOSTIC CODE HISTORY REVEALED THE VENTILATOR'S BACKUP BATTERY HAD DEPLETED DURING USE PRIOR TO THE VENT INOP OCCURRENCE. THE BIOMEDICAL ENGINEER CHARGED THE BACKUP BATTERY OVERNIGHT AND TESTED IT THE NEXT DAY AND IT FAILED TESTING. THE BIOMEDICAL ENGINEER REPLACED THE BACKUP BATTERY. THIS EVENT IS BEING REPORTED AS A USER ERROR. THE CUSTOMER REPORTED THE HOSPITAL WAS EXPERIENCING POWER ISSUES WHICH, WHEN OCCURRING, WILL OPERATE THE VENTILATOR VIA BACKUP BATTERY POWER UNTIL AC POWER IS RESTORED BY THE USER. THE DIAGNOSTIC LOG SHOWED THE VENTILATOR HAD BEEN OPERATING ON BACKUP BATTERY POWER WHICH THEN DEPLETED DURING EXTENDED USE. PER THE VENTILATOR'S OPERATORS MANUAL, WHEN THE VENTILATOR IS POWERED BY BACKUP BATTERY IT WILL GENERATE A NONRESECTABLE, NON SILENCEABLE ALARM EVERY 60 SECONDS. DURING THIS STATE A YELLOW BATTERY IN USE LED IS LIT. OPERATION WILL CONTINUE IN THIS STATE UNTIL THE BATTERY CAPACITY IS NEARLY EXPENDED. WHEN THE BATTERY HAS ONLY ABOUT 5 MINUTES OF OPERATION LEFT, AN AUDIBLE, NONRESECTABLE HIGH PRIORITY ALARM WILL SOUND AND A RED BATTERY LOW LED WILL FLASH CONTINUOUSLY. WHEN THE BATTERY LOW INDICATOR IS FLASHING RED, OPERATION OF THE VENTILATOR FROM THE BATTERY POWER SHOULD BE DISCONTINUED. THE BACKUP BATTERY WILL CHARGE WHEN THE VENTILATOR IS PLUGGED INTO A VIABLE AC SUPPLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ESPRIT VENTILATOR | VENTILATOR, CONTINUOUS | CBK | RESPIRONICS CALIFORNIA, INC. | V1000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |