FDA Adverse Event Malfunction Summary report: N

ATELLICA IM ANTI-HEPATITIS B SURFACE ANTIGEN 2 (AHBS2)

MDR report key: 16448552 · Received February 27, 2023

Report

Report Number
1219913-2023-00039
Event Type
Malfunction
Date Received
February 27, 2023
Date of Event
February 13, 2023
Report Date
May 11, 2023
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS, INC.
Product Code
LOM
UDI-DI
00630414597836
PMA / PMN Number
P100039-S005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A CUSTOMER FROM OUTSIDE THE UNITED STATES REPORTED OBSERVATION OF POSITIVE ATELLICA IM ANTI-HEPATITIS B SURFACE ANTIGEN 2 (AHBS2) RESULTS FOR ONE PATIENT WHICH WERE DISCORDANT RELATIVE TO ALTERNATE-METHOD TESTING. ON (B)(6) 2023, A POSITIVE AHBS2 RESULT OF 27 MIU/ML WAS OBTAINED FOR A FEMALE CHRONIC HEPATITIS PATIENT WHO HAS BEEN UNDERGOING ANTIVIRAL TREATMENT. THIS INITIAL RESULT WAS NOT REPORTED TO THE PHYSICIAN. ADDITIONAL TESTING WAS PERFORMED ON THIS SAMPLE, USING ALTERNATE METHODS AT AN EXTERNAL LAB: HBS ANTIBODIES: <2 (NEGATIVE) - THIS RESULT WAS COMMUNICATED TO THE DOCTOR, AS CONSISTENT WITH CLINICAL EXPECTATION. HBS ANTIGENS: POSITIVE. ON (B)(6) 2023, A RETAINED SAMPLE (DRAWN ON (B)(6) 2023) FROM THE SAME PATIENT WAS SUBJECTED TO ADDITIONAL TESTING: ATELLICA AHBS2: 40.94 MIU/ML (POSITIVE). PCR VIREMIA: NEGATIVE. IT IS NOTED THAT THE SAME PATIENT HAD TESTED NEGATIVE FOR AHBS2 IN 2015, AND THAT THE PATIENT HAD BEEN HBCT AND HBSII POSITIVE SINCE 2013. THE ATELLICA IM TNIH INSTRUCTIONS FOR USE (IFU) STATES THE FOLLOWING, UNDER LIMITATIONS: "RESULTS OF THIS ASSAY SHOULD ALWAYS BE INTERPRETED IN CONJUNCTION WITH PATIENT'S MEDICAL HISTORY, CLINICAL PRESENTATION AND OTHER FINDINGS". SIEMENS IS INVESTIGATING. NOTE: THE DISCORDANT OBSERVATION OF (B)(6) 2023 IS REPORTED IN MDR 1219913-2023-00040.

Additional Manufacturer Narrative · 0

MDR 1219913-2023-00039 WAS INITIALLY SUBMITTED ON 2023-02-27. UPDATE, 2023-04-27: SIEMENS HAS CONCLUDED THE INVESTIGATION. A CUSTOMER FROM OUTSIDE THE UNITED STATES REPORTED OBSERVATION OF POSITIVE ATELLICA IM ANTI-HEPATITIS B SURFACE ANTIGEN 2 (AHBS2) RESULTS FOR ONE PATIENT WHICH WERE DISCORDANT RELATIVE TO ALTERNATE-METHOD TESTING. SIEMENS REQUESTED THE AFFECTED SAMPLE FOR ADDITIONAL TESTING, BUT THERE WAS INSUFFICIENT VOLUME AVAILABLE. IT WAS NOTED THAT THE PATIENT HAS A HISTORY OF CHRONIC HEPATITIS WITH ONGOING ANTIVIRAL TREATMENTS. QUALITY CONTROL (QC) PERFORMANCE WAS REVIEWED, WITH NO ISSUES NOTED. BASED ON THE OBSERVED REPLICATION OF ELEVATED RESULTS, SIEMENS CANNOT RULE OUT PRE-ANALYTICAL FACTORS OR A SAMPLE-SPECIFIC INTERFERENT AS THE CAUSE OF THE REACTIVE TESTING. THE CLINICAL SENSITIVITY AND SPECIFICITY SECTION OF THE ATELLICA IM ANTI-HEPATITIS B SURFACE ANTIGEN 2 (AHBS2) INSTRUCTIONS FOR USE (IFU) STATES THE 95% CONFIDENCE INTERVAL (CI) FOR RESOLVED RELATIVE SPECIFICITY AS 98.34% - 99.88%, SO A CERTAIN NUMBER OF FALSE POSITIVE RESULTS CAN BE EXPECTED FOR THIS ASSAY. THE ASSAY IS PERFORMING AS EXPECTED AND WITHIN SPECIFICATIONS. RESULTS OF THIS ASSAY SHOULD ALWAYS BE INTERPRETED IN CONJUNCTION WITH THE PATIENT'S MEDICAL HISTORY, CLINICAL PRESENTATION, AND OTHER FINDINGS. BASED ON THE INVESTIGATION, NO PRODUCT PROBLEM WAS IDENTIFIED. THE CUSTOMER IS OPERATIONAL AND REQUIRES NO FURTHER SUPPORT. NOTES: 1) IN SECTION H6, THE CODES FOR INVESTIGATION FINDINGS AND INVESTIGATION CONCLUSION HAVE BEEN UPDATED. 2) MDR 1219913-2023-00040 SUPPLEMENTAL 1 HAS BEEN FILED IN ASSOCIATION WITH THE RELATED OBSERVATION OF 2023-02-16.

Description of Event or Problem · 0

THE CUSTOMER REPORTED OBSERVATION OF A POSITIVE ATELLICA IM ANTI-HEPATITIS B SURFACE ANTIGEN 2 (AHBS2) RESULT FOR ONE PATIENT WHICH WAS DISCORDANT RELATIVE TO ALTERNATE-METHOD TESTING. THE NEGATIVE (ALTERNATE-METHOD) RESULT WAS ACCEPTED AS CONSISTENT WITH PATIENT CLINICAL INDICATIONS. THERE ARE NO ALLEGATIONS OF PATIENT HARM, CHANGES IN TREATMENT, OR DELAYS OF DIAGNOSIS IN ASSOCIATION WITH THE OBSERVED DISCORDANCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
384936 ATELLICA IM ANTI-HEPATITIS B SURFACE ANTIGEN 2 (AHBS2) AHBS IMMUNOASSAY LOM SIEMENS HEALTHCARE DIAGNOSTICS, INC. N/A 155 00630414597836

Patients

Seq Age Sex Outcome Treatment
1 Female