EMMA CAPNOGRAPH, KPA
Report
- Report Number
- 3019388613-2023-00046
- Event Type
- Malfunction
- Date Received
- February 27, 2023
- Date of Event
- January 1, 2023
- Report Date
- February 3, 2023
- Manufacturer
- MASIMO - 15750 ALTON PKWY
- Product Code
- CCK
- UDI-DI
- 07350046131040
- PMA / PMN Number
- K063167
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
ATTEMPTS HAVE BEEN MADE TO OBTAIN THE PRODUCT. THE PRODUCT HAS NOT BEEN RETURNED TO MASIMO TO ALLOW AN ANALYSIS TO BE PERFORMED. IF THE PRODUCT IS RETURNED FOR EVALUATION OR NEW INFORMATION IS OBTAINED, A FOLLOW UP REPORT WILL BE SUBMITTED. INITIAL REPORTER ZIP CODE EXCEEDED THE MAXIMUM ALLOWABLE CHARACTERS, ZIP CODE IS AS FOLLOWS: (B)(6). INITIAL REPORTER PHONE NUMBER EXCEEDED MAXIMUM ALLOWABLE DIGITS, PHONE NUMBER IS AS FOLLOWS: (B)(6).
ADDITIONAL MANUFACURING NARRATIVE: OTHER, OTHER TEXT: THE RETURNED DEVICE WAS EVALUATED. THE EMMA MEASURED OUTSIDE OF THE SPECIFICATION RANGE FOR CO2 MEASUREMENTS BECAUSE IT WAS NOT ZEROED OUT AS PART OF THE PREVENTATIVE MAINTENANCE. ONCE THE ZEROING PROCEDURE WAS COMPLETED, THE MEASURED VALUE WAS WITHIN THE SPECIFIED RANGE. INITIAL REPORTER ZIP CODE EXCEEDED THE MAXIMUM ALLOWABLE CHARACTERS, ZIP CODE IS AS FOLLOWS: (B)(6). INITIAL REPORTER PHONE NUMBER EXCEEDED MAXIMUM ALLOWABLE DIGITS, PHONE NUMBER IS AS FOLLOWS: (B)(6).
THE CUSTOMER REPORTED THE DEVICE SHOWING LOW READING ON 5% CO2. NO CONSEQUENCES OR IMPACT TO PATIENT WERE REPORTED.
THE CUSTOMER REPORTED THE DEVICE SHOWING LOW READING ON 5% CO2. NO CONSEQUENCES OR IMPACT TO PATIENT WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 787348 | EMMA CAPNOGRAPH, KPA | ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE | CCK | MASIMO - 15750 ALTON PKWY | 18266 | 07350046131040 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |