FDA Adverse Event Malfunction Summary report: N

EMMA CAPNOGRAPH, KPA

MDR report key: 16445706 · Received February 27, 2023

Report

Report Number
3019388613-2023-00046
Event Type
Malfunction
Date Received
February 27, 2023
Date of Event
January 1, 2023
Report Date
February 3, 2023
Manufacturer
MASIMO - 15750 ALTON PKWY
Product Code
CCK
UDI-DI
07350046131040
PMA / PMN Number
K063167
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ATTEMPTS HAVE BEEN MADE TO OBTAIN THE PRODUCT. THE PRODUCT HAS NOT BEEN RETURNED TO MASIMO TO ALLOW AN ANALYSIS TO BE PERFORMED. IF THE PRODUCT IS RETURNED FOR EVALUATION OR NEW INFORMATION IS OBTAINED, A FOLLOW UP REPORT WILL BE SUBMITTED. INITIAL REPORTER ZIP CODE EXCEEDED THE MAXIMUM ALLOWABLE CHARACTERS, ZIP CODE IS AS FOLLOWS: (B)(6). INITIAL REPORTER PHONE NUMBER EXCEEDED MAXIMUM ALLOWABLE DIGITS, PHONE NUMBER IS AS FOLLOWS: (B)(6).

Additional Manufacturer Narrative · 0

ADDITIONAL MANUFACURING NARRATIVE: OTHER, OTHER TEXT: THE RETURNED DEVICE WAS EVALUATED. THE EMMA MEASURED OUTSIDE OF THE SPECIFICATION RANGE FOR CO2 MEASUREMENTS BECAUSE IT WAS NOT ZEROED OUT AS PART OF THE PREVENTATIVE MAINTENANCE. ONCE THE ZEROING PROCEDURE WAS COMPLETED, THE MEASURED VALUE WAS WITHIN THE SPECIFIED RANGE. INITIAL REPORTER ZIP CODE EXCEEDED THE MAXIMUM ALLOWABLE CHARACTERS, ZIP CODE IS AS FOLLOWS: (B)(6). INITIAL REPORTER PHONE NUMBER EXCEEDED MAXIMUM ALLOWABLE DIGITS, PHONE NUMBER IS AS FOLLOWS: (B)(6).

Description of Event or Problem · 0

THE CUSTOMER REPORTED THE DEVICE SHOWING LOW READING ON 5% CO2. NO CONSEQUENCES OR IMPACT TO PATIENT WERE REPORTED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THE DEVICE SHOWING LOW READING ON 5% CO2. NO CONSEQUENCES OR IMPACT TO PATIENT WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
787348 EMMA CAPNOGRAPH, KPA ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE CCK MASIMO - 15750 ALTON PKWY 18266 07350046131040

Patients

Seq Age Sex Outcome Treatment
1 Unknown