FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK¿ SYRINGE STERILE

MDR report key: 16433759 · Received February 24, 2023

Report

Report Number
9614033-2023-00011
Event Type
Malfunction
Date Received
February 24, 2023
Date of Event
January 13, 2023
Report Date
June 6, 2023
Manufacturer
BECTON DICKINSON DE MEXICO
Product Code
FMF
UDI-DI
00382903033102
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DC, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: DEVICE AVAILABLE FOR EVAL YES, RETURNED TO MANUFACTURER ON: 07-MAR-2023 INVESTIGATION SUMMARY TWO SAMPLES AND PHOTOS RECEIVED BY OUR QUALITY TEAM FOR INVESTIGATION. UPON VISUAL EVALUATION, CRACKED BARREL ON SYRINGE IS OBSERVED, WHICH LIKELY CAUSED THE LEAKAGE. A DEVICE HISTORY REVIEW WAS PERFORMED FOR LOT 2221243 NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED ISSUE. BASED ON THE TEAMS INVESTIGATION, POSSIBLE ROOT CAUSE IS ASSOCIATED WITH MISALIGNMENT BETWEEN THE INPUT AND TRANSFER DISKS.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

MIT WAS REPORTED THAT 2 BD LUER-LOK¿ SYRINGE STERILE WAS DAMAGED AND LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: 22G IV PLACED IN PT'S AC. RN ATTACHED 20ML BD SYRINGE LOT #2221243 TO END OF IV CONNECTOR. WITHOUT PULLING BACK, BLOOD FLOWED FROM IV INTO SYRINGE AND THE SYRINGE CRACKED ALONG THE BARREL. RN TRIED A SECOND 20ML SYRINGE WITH THE SAME RESULT. OF NOTE, OTHER SYRINGES WITH THIS LOT NUMBER WERE USED ON THIS PATIENT WITH NO ISSUES. SYRINGES WERE SAVED AND WE WOULD LIKE TO SEND TO BD FOR EVALUATION AND REPORT.

Description of Event or Problem · 0

MIT WAS REPORTED THAT 2 BD LUER-LOK¿ SYRINGE STERILE WAS DAMAGED AND LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: 22G IV PLACED IN PT'S AC. RN ATTACHED 20ML BD SYRINGE LOT #2221243 TO END OF IV CONNECTOR. WITHOUT PULLING BACK, BLOOD FLOWED FROM IV INTO SYRINGE AND THE SYRINGE CRACKED ALONG THE BARREL. RN TRIED A SECOND 20ML SYRINGE WITH THE SAME RESULT. OF NOTE, OTHER SYRINGES WITH THIS LOT NUMBER WERE USED ON THIS PATIENT WITH NO ISSUES. SYRINGES WERE SAVED AND WE WOULD LIKE TO SEND TO BD FOR EVALUATION AND REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1934400 BD LUER-LOK¿ SYRINGE STERILE PISTON SYRINGE FMF BECTON DICKINSON DE MEXICO 2221243 00382903033102

Patients

Seq Age Sex Outcome Treatment
1 Unknown