FDA Adverse Event Malfunction Summary report: N

KIT BD MAX CHECK-POINTS CPO IVD

MDR report key: 16429471 · Received February 23, 2023

Report

Report Number
1119779-2023-00171
Event Type
Malfunction
Date Received
February 23, 2023
Date of Event
January 19, 2023
Report Date
March 24, 2023
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
POC
PMA / PMN Number
K190613
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING INFORMATION HAS BEEN UPDATED; B.5. EVENT DESCRIPTION UPDATED TO INCLUDE THE CORRECT BRAND NAME. KIT BD MAX CHECK-POINTS CPO IVD. D.1. MEDICAL DEVICE BRAND NAME: KIT BD MAX CHECK-POINTS CPO IVD. D.4. MEDICAL DEVICE CATALOG #: 278102. D.4. MEDICAL DEVICE LOT # : 2017254. D.4. MEDICAL DEVICE EXPIRATION DATE: 10/07/2022. D.4. UNIQUE IDENTIFIER (UDI) #: (B)(4). AFTER FURTHER EVALUATION OF THE COMPLAINT, IT HAS BEEN DETERMINED THAT THE PREVIOUSLY SUBMITTED REPORT 1119779-2023-00171 WAS SENT IN ERROR. THE PRODUCT INVOLVED IS NOT SOLD WITHIN THE US AND BD DOES NOT SELL A SIMILAR PRODUCT WITHIN THE US AND/OR IT IS NOT A REGISTERED MEDICAL DEVICE IN THE US. THIS COMPLAINT IS NOT MDR REPORTABLE.

Additional Manufacturer Narrative · 0

COMMON DEVICE NAME: SYSTEM, NUCLEIC ACID AMPLIFICATION TEST, DNA, ANTIMICROBIAL RESISTANCE MARKER, DIRECT SPECIMEN. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT USING KIT BD MAX CHECK-POINTS CPO IVD FALSE NEGATIVES HAVE OCCURRED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MAX CT0488 A5, RUN 2922 - NDM AND VIM POSITIVE. NGS STRAINS CULTURED POSITIVE: KPC, OXA-48, NDM, VIM. FALSE NEGATIVE: KPC AND OXA-48. MAX CT0488 B2, RUN 2933 - ALL NEGATIVE. NGS STRAINS CULTURED POSITIVE: VIM AND OXA-48. FALSE NEGATIVE: VIM AND OXA-48.

Description of Event or Problem · 0

IT WAS REPORTED THAT KIT BD MAX CPO FALSE NEGATIVES HAVE OCCURRED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MAX CT0488 A5, RUN 2922 - NDM AND VIM POSITIVE. NGS STRAINS CULTURED POSITIVE: KPC, OXA-48, NDM, VIM. FALSE NEGATIVE: KPC AND OXA-48. MAX CT0488 B2, RUN 2933 - ALL NEGATIVE. NGS STRAINS CULTURED POSITIVE: VIM AND OXA-48. FALSE NEGATIVE: VIM AND OXA-48.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
678378 KIT BD MAX CHECK-POINTS CPO IVD SEE H.10 POC BECTON, DICKINSON & CO. (SPARKS) 2017254

Patients

Seq Age Sex Outcome Treatment
1 Unknown