TFNA FEM NAIL Ø11 R 130° L360 TIMO15
Report
- Report Number
- 8030965-2023-02227
- Event Type
- Injury
- Date Received
- February 23, 2023
- Date of Event
- February 2, 2023
- Manufacturer
- SYNTHES GMBH
- Product Code
- HSB
- UDI-DI
- 07611819650602
- PMA / PMN Number
- K160167
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. DEVICE AVAILABLE FOR EVALUATION: COMPLAINANT PART IS NOT EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION. THE INITIAL REPORTER IS A J&J SALES REPRESENTATIVE. PMA/510(K): K131548. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. 11. CORRECTED DATA: H4 DEVICE MANUFACTURE DATE CORRECTED. MANUFACTURING LOCATION: MONUMENT. MANUFACTURING DATE: 16-JUL-2021. EXPIRATION DATE: 01-JUL-2031. PART NUMBER: 04.037.156S, 11MM/130 DEG TI CANN TFNA 360MM/RIGHT- STERILE. LOT NUMBER: 268P688 (STERILE). PRODUCTION ORDER TRAVELER MET ALL INSPECTION ACCEPTANCE CRITERIA. INSPECTION SHEET, IN-PROCESS / INSPECT DIMENSIONAL / FINAL MET ALL INSPECTION ACCEPTANCE CRITERIA. INSPECTION SHEET, TFNA ASSEMBLY INSPECTION MET ALL INSPECTION ACCEPTANCE CRITERIA. PACKAGING LABEL LOG (PLL) LPPF REV E, LMD REV A WAS REVIEWED AND DETERMINED TO BE CONFORMING. PACKAGING BOM WAS REVIEWED AND DETERMINED TO BE CONFORMING WITH NO DEVIATIONS TO NORMAL PACKAGING IDENTIFIED. SCN SUPPLIED BY ETHICON (ABQ) WAS REVIEWED AND DETERMINED TO BE CONFORMING. THIS LOT MET ALL DIMENSIONAL, VISUAL, STERILITY AND PACKAGING CRITERIA AT THE TIME OF RELEASE WITH NO ISSUES DOCUMENTED DURING THE MANUFACTURE THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. COMPONENT PART(S) REVIEWED: COMPONENT PARTS WERE NOT REVIEWED AS THE REPORTED COMPLAINT CONDITION OF ¿REVISED WITH ANGLED BLADE PLATE DUE TO NON-UNION¿ DOES NOT INDICATE BREAKAGE OF THE NAIL OR ANY OF ITS COMPONENTS. THEREFORE, REVIEW OF THE RAW MATERIALS WOULD NOT BE PERTINENT TO THE REPORTED COMPLAINT CONDITION. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN UNITED KINGDOM AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THE PATIENT WAS HAVING A FAILED TROCHANTERIC FIXATION NAIL ADVANCED (TFNA) NAIL WAS REMOVED IN WHICH THE FRACTURE HAD FAILED TO UNITE. THE ORIGINAL TFNA WAS IMPLANTED ON (B)(6) 2021 AND WAS REVISED WITH AN ANGLED BLADE PLATE. REVISION PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS REPORT IS FOR ONE (1) 11MM/130 DEG TI CANN TFNA 360MM/RIGHT - STERILE. THIS IS REPORT 1 OF 3 FOR COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 594364 | TFNA FEM NAIL Ø11 R 130° L360 TIMO15 | ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES | HSB | SYNTHES GMBH | 268P688 | 07611819650602 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Required Intervention | TFNA SCR PERF L95 TAN.| UNK - SCREWS: LOCKING. |