FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 2.0 TEST KIT 24T (EUA)

MDR report key: 16419193 · Received February 22, 2023

Report

Report Number
1221359-2023-00389
Event Type
Malfunction
Date Received
February 22, 2023
Date of Event
February 9, 2023
Report Date
May 12, 2023
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
UDI-DI
00811877011354
PMA / PMN Number
EUA210517
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED IN DUPLICATE AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 1100027 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR KIT PART NUMBER 192-000 / LOT 1100027 AND TEST BASE PART NUMBER 192-430 / LOT 1100027. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. THE REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED, UNCONFIRMED AND CONFLICTING RESULTS) RELATED TO KIT LOT 1100027 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, A POSSIBLE ASSIGNABLE ROOT CAUSE IS PATIENT SAMPLE INTERFERENCE. SUBSTANCES IN THE SAMPLE ITSELF MAY HAVE INTERFERED WITH THE REACTION. H3 OTHER TEXT : DEVICE DISCARDED; SINGLE-USE DEVICE.

Additional Manufacturer Narrative · 0

THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. DEVICE DISCARDED; SINGLE-USE DEVICE.

Description of Event or Problem · 0

THE CUSTOMER REPORTED A FALSE NEGATIVE RESULT WITH ID NOW COVID-19 2.0 ASSAY PERFORMED ON (B)(6) 2023 WITH NASAL SWAB SAMPLE. CONFIRMATION TESTING ON NASOPHARYNGEAL SWAB IN TRANSPORT MEDIA VIA PCR (PLATFORM DIASORIN) WAS PERFORMED AT (B)(6) ON (B)(6) 2023 AND GENERATED A POSITIVE RESULT. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED A FALSE NEGATIVE RESULT WITH ID NOW COVID-19 2.0 ASSAY PERFORMED ON (B)(6) 2023 WITH NASAL SWAB SAMPLE. CONFIRMATION TESTING ON NASOPHARYNGEAL SWAB IN TRANSPORT MEDIA VIA PCR (PLATFORM DIASORIN) WAS PERFORMED AT (B)(6) ON (B)(6) 2023 AND GENERATED A POSITIVE RESULT. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1319297 ID NOW COVID-19 2.0 TEST KIT 24T (EUA) REAGENTS, 2019-NOVEL CORONAVIRUS NUCLEIC ACID QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 1100027 00811877011354

Patients

Seq Age Sex Outcome Treatment
1 Unknown