MENTOR SMOOTH ROUND MODERATE PLUS PROFILE
Report
- Report Number
- 1645337-2023-01942
- Event Type
- Injury
- Date Received
- February 22, 2023
- Date of Event
- November 1, 2021
- Manufacturer
- MENTOR TEXAS
- Product Code
- FWM
- UDI-DI
- 00081317001423
- PMA / PMN Number
- P990075
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
ON FEBRUARY 24, 2023, MENTOR BECAME AWARE REGARDING THE IMPACTED DEVICE INFORMATION. HENCE, THE FOLLOWING FIELDS HAVE BEEN UPDATED ON THIS FORM: FIELD D1 FOR BRAND NAME HAS BEEN UPDATED TO "MENTOR SMOOTH ROUND MODERATE PLUS PROFILE" REPLACEMENT DEVICE USED ON (B)(6), 2023 WAS SERIAL NUMBER (B)(6) ON THE LEFT SIDE, AND SERIAL NUMBER (B)(6) ON THE RIGHT SIDE FIELD D4 FOR CATALOG NUMBER HAS BEEN UPDATED TO "3502375" FIELD D4 FOR LOT NUMBER HAS BEEN UPDATED TO "5820142" FIELD D4 FOR UNIQUE IDENTIFIER( UDI) HAS BEEN UPDATED TO "(B)(4)." FIELD G4 FOR PMA/ 510(K) HAS BEEN UPDATED TO "P990075" A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. THERE IS AN INCONSISTENCY BETWEEN THE DATE OF IMPLANT AND THE DATE OF MANUFACTURE, HOWEVER, THE PATIENT ONLY REMEMBERS THAT IT WAS 2008, IF ADDITIONAL INFORMATION IS RECEIVED IN THE FUTURE, IT WILL BE DOCUMENTED AND SUPPLEMENTAL REPORT WILL BE SUBMITTED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
ON (B)(6) 2023, DEVICE EVALUATION WAS COMPLETED. DEVICE EVALUATION SUMMARY: MENTOR CONDUCTED A VISUAL INSPECTION, LEAK TESTING, AND MICROSCOPIC EXAMINATION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE SM MPPS DV 375CC BREAST IMPLANT HAD A CREASE/FOLD ON THE ANTERIOR VIEW EXTENDING TO THE POSTERIOR VIEW. LEAK TESTING WAS PERFORMED, ACCORDING TO MENTOR PROCEDURE, AND IT IDENTIFIED A TEAR ON THE ANTERIOR VIEW WITHIN THE CREASE/FOLD MEASURING APPROXIMATELY 0.1 CM. A MICROSCOPIC EXAMINATION WAS PERFORMED AND INSTRUMENT DAMAGE WAS NOT FOUND ON THE AREA OF THE TEAR. THE EVALUATION DETERMINED THAT THE POSSIBLE CAUSE OF THE RUPTURE WAS CONSISTENT WITH NORMAL WEAR. DEFLATION CAN OCCUR AT ANY TIME AFTER IMPLANTATION, BUT IT IS MORE LIKELY TO OCCUR THE LONGER THE IMPLANT IS IMPLANTED. CREATING WRINKLES OR FOLDS SHOULD BE AVOIDED DURING IMPLANTATION OR OTHER PROCEDURES AS IT CAN RESULT IN DEFLATION AT A LATER TIME. BREAST IMPLANTS MAY ALSO SIMPLY WEAR OUT OVER TIME. A SECOND PRODUCT WAS RECEIVED (LOT-5824256). NO ADVERSE EVENTS WERE REPORTED FOR THIS CONCOMITANT (CONTRALATERAL) DEVICE, THEREFORE NO FURTHER ANALYSIS IS REQUIRED. AS PART OF MENTOR¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. CAPA ACTIVITY WAS NOT REQUIRED. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST-MARKET SURVEILLANCE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. SINCE NO LOT NUMBER WAS PROVIDED, NO MANUFACTURING RECORD EVALUATION COULD BE PERFORMED. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: LEFT DEFLATION MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
ON MARCH 30, 2023, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
IT WAS REPORTED THAT A 36-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT BREAST AUGMENTATION PROCEDURE WITH UNKNOWN SIZE UNKNOWN SALINE IMPLANTS AND EXPERIENCED BREAST IMPLANT DEFLATION ON HER LEFT SIDE POSTOPERATIVELY; CONFIRMED VIA ULTRASOUND. THE PATIENT HAS CONSULTED WITH THE HEALTHCARE PROFESSIONAL. AS A RESULT, THE DEVICE WAS EXPLANTED ON (B)(6) 2023.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1642887 | MENTOR SMOOTH ROUND MODERATE PLUS PROFILE | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | MENTOR TEXAS | 3502375 | 5820142 | 00081317001423 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 36 YR | Female | Required Intervention |