INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE
Report
- Report Number
- 1037905-2023-00084
- Event Type
- Injury
- Date Received
- February 22, 2023
- Date of Event
- January 29, 2023
- Report Date
- May 3, 2023
- Manufacturer
- WILSON-COOK MEDICAL INC
- Product Code
- PKL
- UDI-DI
- 00827002580107
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
PMA/510(K): K212323. INVESTIGATION EVALUATION: THE PRODUCTS SAID TO BE INVOLVED WERE RETURNED, 3 FROM THE LOT NUMBER PROVIDED IN THIS REPORT. DEVICE #1-3: OUR LABORATORY EVALUATION OF THE PRODUCT SAID TO BE INVOLVED COULD NOT CONFIRM THE REPORT AS IT WAS DESCRIBED, BECAUSE ALL THE DEVICE COMPONENTS (PARTICULARLY THE CLIP) WERE NOT INCLUDED IN THE RETURN. DURING FUNCTIONAL TESTING THE DEVICE WAS ADVANCED INTO THE ACCESSORY CHANNEL OF AN OLYMPUS GASTROSCOPE (2.8MM CHANNEL, GIF-020) WHICH WAS PLACED IN A SIMULATED UPPER GI POSITION. THE TIP OF THE ENDOSCOPE WAS RETROFLEXED TO SIMULATE WORST CASE SCENARIO. WITH HANDLE MANIPULATION, THE DRIVE WIRE WAS OBSERVED TO MOVE FREELY INSIDE THE OUTER SHEATH. A VISUAL EXAMINATION OF THE DRIVE WIRE, CATHETER ATTACHMENT, AND COIL CATHETER (DISTAL END DEVICE COMPONENTS) SHOWED NO DEFORMITIES OR SIGNS OF DAMAGE. A PRODUCT DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THE REPORTED OCCURRENCE WAS NOT OBSERVED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: THE ADDITIONAL INFORMATION INDICATED THE USER HELD THE HANDLE SPOOL DURING ADVANCEMENT THROUGH THE ACCESSORY CHANNEL. THIS IS THE MOST LIKELY CAUSE OF THIS REPORT. THE INSTRUCTIONS FOR USE STATES: "WITH CLIP CLOSED AND WITHOUT HOLDING HANDLE SPOOL, ADVANCE DEVICE IN SMALL INCREMENTS INTO ACCESSORY CHANNEL OF GASTROSCOPE OR COLONOSCOPE. CAUTION: HOLDING HANDLE SPOOL DURING CLIP ADVANCEMENT MAY PREMATURELY DEPLOY CLIP." FAILURE TO FOLLOW THE INSTRUCTIONS ABOVE CAN RESULT IN DAMAGE TO THE DEVICE WHICH MAY LEAD TO PREMATURE CLIP DEPLOYMENT. HOWEVER, EVEN IF THE CLIP IS NOT DEPLOYED, DAMAGE CAN OCCUR TO INTERNAL DEVICE COMPONENTS SUCH AS THE DRIVER LEGS. ONCE THIS DAMAGE OCCURS, THE CLIP IS IN A PARTIALLY DEPLOYED STATE AND MAY NOT BE ABLE TO BE OPENED/REOPENED. PRIOR TO DISTRIBUTION, ALL INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A REVIEW OF THE COMPLAINT HISTORY WAS CONDUCTED. THE LIKELIHOOD OF OCCURRENCE IS CONSIDERED RARE. CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE. ADDITIONAL COMMENTS REGARDING THIS REPORT: BASED ON THE INFORMATION PROVIDED THAT THE USER HELD THE HANDLE SPOOL DURING ADVANCEMENT THROUGH THE ACCESSORY CHANNEL, A COOK REPRESENTATIVE HAS CONTACTED THE MEDICAL FACILITY INVOLVED IN AN EFFORT TO PROMOTE FURTHER EDUCATION AND UNDERSTANDING RELATED TO APPROPRIATE USAGE OF THIS PRODUCT.
PMA/510(K): K212323. THE INVESTIGATION IS ONGOING. A FOLLOW-UP EMDR WILL BE PROVIDED WITHIN 30 DAYS OF SUBMISSION OF THIS REPORT.
PMA/510(K): K212323. INVESTIGATION EVALUATION: A PRODUCT EVALUATION WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVALUATION. THE REPORT COULD NOT BE CONFIRMED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: A CORRECTIVE ACTION HAS BEEN INITIATED TO REDUCE OCCURRENCES OF THE CLIP HOUSING DETACHING FROM THE CATH ATTACH FOR INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICES. THE PRODUCT SAID TO BE INVOLVED IS INCLUDED IN THE SCOPE OF THE CORRECTIVE ACTIONS. THE INSTRUCTIONS FOR USE STATES: "UNCOIL DEVICE. VERIFY SMOOTH HANDLE OPERATION AND CLIP ACTION. OPEN CLIP BY GENTLY MOVING HANDLE SPOOL DISTALLY (AWAY FROM HANDLE THUMB RING). ONCE CLIP IS FULLY OPEN, DO NOT CONTINUE ADVANCING HANDLE SPOOL AS CLIP MAY PREMATURELY DETACH FROM CATHETER. CLOSE CLIP BY MOVING HANDLE SPOOL PROXIMALLY UNTIL CLIP IS FULLY CLOSED. PRECAUTION: DO NOT CONTINUE TO PULL HANDLE SPOOL BEYOND TACTILE RESISTANCE AS THIS MAY PREMATURELY DEPLOY CLIP." THE INSTRUCTIONS FOR USE STATES: "WITH CLIP CLOSED AND WITHOUT HOLDING HANDLE SPOOL, ADVANCE DEVICE IN SMALL INCREMENTS INTO ACCESSORY CHANNEL OF GASTROSCOPE, DUODENOSCOPE, OR COLONOSCOPE. NOTE: IF DIFFICULT TO ADVANCE DEVICE, RELAX ELEVATOR AND/OR STRAIGHTEN ENDOSCOPE.¿ THE IFU ALSO INCLUDES THE FOLLOWING PRECAUTION: ¿HOLDING HANDLE SPOOL DURING CLIP ADVANCEMENT MAY PREMATURELY DEPLOY CLIP.¿ FAILURE TO FOLLOW THE INSTRUCTIONS ABOVE CAN RESULT IN DAMAGE TO THE DEVICE WHICH MAY LEAD TO PREMATURE CLIP DEPLOYMENT. PRIOR TO DISTRIBUTION, ALL INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN INITIATED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. THIS PRODUCT WAS MANUFACTURED PRIOR TO IMPLEMENTATION OF THIS CORRECTIVE ACTION. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.
PMA/510(K): (B)(4). INVESTIGATION EVALUATION: THE PRODUCTS SAID TO BE INVOLVED WERE RETURNED, 3 FROM THE LOT NUMBER PROVIDED IN THIS REPORT. DEVICE #1-3: OUR LABORATORY EVALUATION OF THE PRODUCT SAID TO BE INVOLVED COULD NOT CONFIRM THE REPORT AS IT WAS DESCRIBED, BECAUSE ALL THE DEVICE COMPONENTS (PARTICULARLY THE CLIP) WERE NOT INCLUDED IN THE RETURN. DURING FUNCTIONAL TESTING THE DEVICE WAS ADVANCED INTO THE ACCESSORY CHANNEL OF AN OLYMPUS GASTROSCOPE (2.8MM CHANNEL, GIF-020) WHICH WAS PLACED IN A SIMULATED UPPER GI POSITION. THE TIP OF THE ENDOSCOPE WAS RETROFLEXED TO SIMULATE WORST CASE SCENARIO. WITH HANDLE MANIPULATION, THE DRIVE WIRE WAS OBSERVED TO MOVE FREELY INSIDE THE OUTER SHEATH. A VISUAL EXAMINATION OF THE DRIVE WIRE, CATHETER ATTACHMENT, AND COIL CATHETER (DISTAL END DEVICE COMPONENTS) SHOWED NO DEFORMITIES OR SIGNS OF DAMAGE. A PRODUCT DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THE REPORTED OCCURRENCE WAS NOT OBSERVED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: A CORRECTIVE ACTION HAS BEEN INITIATED TO REDUCE OCCURRENCES OF THE CLIP HOUSING DETACHING FROM THE CATH ATTACH FOR INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICES. THE PRODUCT SAID TO BE INVOLVED IS INCLUDED IN THE SCOPE OF THE CORRECTIVE ACTIONS. THE INSTRUCTIONS FOR USE STATES: "UNCOIL DEVICE. VERIFY SMOOTH HANDLE OPERATION AND CLIP ACTION. OPEN CLIP BY GENTLY MOVING HANDLE SPOOL DISTALLY (AWAY FROM HANDLE THUMB RING). ONCE CLIP IS FULLY OPEN, DO NOT CONTINUE ADVANCING HANDLE SPOOL AS CLIP MAY PREMATURELY DETACH FROM CATHETER. CLOSE CLIP BY MOVING HANDLE SPOOL PROXIMALLY UNTIL CLIP IS FULLY CLOSED. PRECAUTION: DO NOT CONTINUE TO PULL HANDLE SPOOL BEYOND TACTILE RESISTANCE AS THIS MAY PREMATURELY DEPLOY CLIP." THE INSTRUCTIONS FOR USE STATES: "WITH CLIP CLOSED AND WITHOUT HOLDING HANDLE SPOOL, ADVANCE DEVICE IN SMALL INCREMENTS INTO ACCESSORY CHANNEL OF GASTROSCOPE, DUODENOSCOPE, OR COLONOSCOPE. NOTE: IF DIFFICULT TO ADVANCE DEVICE, RELAX ELEVATOR AND/OR STRAIGHTEN ENDOSCOPE.¿ THE IFU ALSO INCLUDES THE FOLLOWING PRECAUTION: ¿HOLDING HANDLE SPOOL DURING CLIP ADVANCEMENT MAY PREMATURELY DEPLOY CLIP.¿ FAILURE TO FOLLOW THE INSTRUCTIONS ABOVE CAN RESULT IN DAMAGE TO THE DEVICE WHICH MAY LEAD TO PREMATURE CLIP DEPLOYMENT. PRIOR TO DISTRIBUTION, ALL INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN INITIATED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. THIS PRODUCT WAS MANUFACTURED PRIOR TO IMPLEMENTATION OF THIS CORRECTIVE ACTION. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.
DURING AN ENDOSCOPIC PROCEDURE FOR ACUTE BLEEDING AT THE BULBUS, THE PHYSICIAN USED A COOK INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE. IT WAS REPORTED THAT WHEN THE DISTAL PART OF THE CLIP APPEARED IN THE ENDOSCOPY IMAGE, ITS TIP PART WAS PARTIALLY DETACHED FROM THE CATHETER. ALTERNATIVELY, THE PATIENT HAD TO BE TREATED WITH CYANOACRYLATE GLUE TREATMENT AND THE TREATMENT WAS DELAYED IN THIS REGARD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1642735 | INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE | PKL, LIGATOR, HEMORRHOIDAL | PKL | WILSON-COOK MEDICAL INC | G58010 | W4598860 | 00827002580107 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Required Intervention | SCOPE UNKNOWN MAKE AND MODEL |